GE Healthcare · CTHiSpeed
System would not boot with PANIC: KERNEL FAULT reported.
HiSpeed CTi 6.X with Octane and HSA Gantry
Symptom
System would not boot with PANIC: KERNEL FAULT reported. System would not boot and monitor screens turn blank with PANIC: KERNEL FAULT.
Error message
Error ALERT: PCI BRIDGE ERROR found in SYSLOG See long description for more details ALERT: PCI BRIDGE ERROR May 3 08:49:14 1A:ct99 unix: ALERT: PCI BRIDGE ERROR: int_status is 0x4000000 for /hw/node/xtalk/15/pci Oct 26 18:16:49 2E:wcth savecore:pb 29: PANIC: PCI Bridge Error interrupt killed the system
Applies to
- Models
- HiSpeed
- Product
- HiSpeed CTi 6.X HiSpeed CTi 6.X with Octane and HSA Gantry HiSpeed CTi 6.X with Octane and HSA Gantry Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2169940-53 OCTANE PCI CHASSIS See the LD for FEMC part comments 2169940-53 OCTANE PCI CHASSIS Part #: 2169940-53 Part Type: GP REPAIRABLES Part Weight: 2.73 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: OCTANE PCI CHASSIS FRU: Yes Status: Orderable FLS - OCTANE 1 AND 2 PCI CHASSIS. PART MAY BE AVAILABLE AS A UU PART. THIS PART IS THE SAME PART AS 2169940-7; HOWEVER, IT IS SOURCED FROM A DIFFERENT SECTION OF SGI THAN THE MFG VERSION. 10/01-COMPONENT COLLECTION AT RDC 5/23/02- C2R COMPLETED. SHIP AS T3202LE. 03F03FW48 - T3202LE REPLACED BY T3202PE T3202EQ is the PCI Chassis with Boards.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required