Pular para o conteúdo principal
GE Healthcare · CTLightSpeed23000006929144370005000324400036000

LightSpeed RT16 & RT16 Xtra and Vision

Symptom

"Scan hardware Stopped Scan" with Taxi data violation errors. "Scan hardware Stopped Scan" with Taxi data violation errors

Error message

Do Not Use - use one of the other Error: 230000069 Two, or more, unexpected events were detected during scan data transfer. The 230000069 Error has several varation see below. Sun Sep 5 11:10:31 2004 Error: 230000069 Host: darc Process: dataacq File: dip_thread.cxx Line: 396 Function: Data Acquisition : Scan Data Acquisition Scan Type: Scout Scan: 4167/1/1 Scan Seq Id: 1000 Two, or more, unexpected events were detected during scan data transfer. Bit Count Event description 3 1 TAXI modem violation 5 1 View length error Mon Dec 15 10:40:11 2008 span style="COLOR: #ff0000"> Error: 230000069 Host: darc Process: dataacq File: hddip.cxx Line: 720 Function: Data Acquisition : Scan Data Acquisition Scan Type: Helical Scan: 29144/9/3 Scan Seq Id: 37000 Two, or more, unexpected events were detected during scan data transfer. Bit Count Event description 10 1 View Data Corruption Detected: ---------------------- Bad View Details: Bit Count Event description 0 1 Checksum Field Error 1 1 View Number Field Error 2 1 Row Number Field Error 3 1 Data Length Field Error 4 1 Data Type Field Error 5 1 Magic Number Field Error 7 1 Consecutive Corrupted Segments Detected Mon Dec 22 07:16:59 2008 Error: 

Applies to

Models
LightSpeed
Product
LightSpeed RT16 & RT16 Xtra and Vision WB_H_G16_G_GTL = RT16, Xtra WB_H_G16_G_GTL650 = Vision WB_H_G16_G_GTM = RT16, Xtra or Vision LightSpeed RT16 & RT16 Xtra with GRE and GDAS16 gantry Links to the Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2333611 4.X-5.X 850 MB RF RECEIVER See Long Description Part #: 2333611 Part Type: GP CONSUMABLES Part Weight: 0.14 LBR Repairable: No Returnable: No Hazardous Material: N Description: 850 MB RECEIVER FRU: Yes Status: Orderable Comments: FLS - 850MB RECEIVER ADDED PER RSPL. USED ON H16. MSA8429 06/26/02
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.