Symptom

Cradle will move a few inches and stop. Cradle will move in a few inches and stop

Error message

Error: 182211 Velocity error detected by hardware (LM628 excessive position error). See the Long Description for the full error Tue Aug 29 08:09:30 2006 Error: 182211 Host: ETC Process: Cradle File: cmcLm628IsrTsk.c 1.19 10/16/9 Line: 185 Function: PATIENT ANATOMY POSITIONING : Cradle Drive Scan: 20525/2/1 Type: Helical Exception Level: Pri/Soft Time: 08:09:30:260 Log Series: 595 HW error: Velocity error detected by hardware (LM628 excessive position error). Expected Velocity 75000 (.001 mm/sec) Actual Velocity 0 (.01 mm/sec) Pos Err Tol 2540 (.01 mm) Cur Pos 90852 (.01 mm) Cradle Dir: 1 (0 = Out 1 = In [Gantry]).

Applies to

Models
HiSpeed
Product
HiSpeed CTi 6.X HiSpeed CTi 6.X with Octane and HSA Gantry HiSpeed CTi 6.X with Octane and HSA Gantry Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2269175-3 LONGITUDINAL ENCODER ASM (Used on HSA, CTi, H1 QXi, Plus, Ultra, 16 slice H4 and HiSpeed QXi.) replaces 2269175-2. See long description for details Part #: 2269175-3 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Longitudinal Encoder Assembly with Dynapar encoder FRU: Yes Status: Orderable Comments: Per ECR 2036924, 212008898, 12/4/2006. ** FE Assist ** If no stock or unable to order the 2269175-3 , order 2269175-2. (212001261 5/22/07) Part #: 2269175-2 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Longitudinal Encoder Asm. NOTE:This asm. has LCC encoder/tested and sensor gap is 0.06-0.08mm FRU: Yes Status: Orderable Comments: Per ECR 2036924, 212008898, 12/4/2006
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.