Pular para o conteúdo principal
GE Healthcare · CTBrightSpeed20050200510989200107023

BrightSpeed 4/8/16 slice HPower Scanner Elite-16, Edge-8, Excel-4

Symptom

Erratic CT ( software ) operation with image allocation failures

Error message

Error: 200502005 eeeee/s/i not found in image management, cannot reallocate image space See Long Description for sample error Tue Oct 19 09:22:20 2004 Error: 200502005 Host: dgct Process: scanRx File: srx_sma_manage_image_space.c Line: 951 Function: No system function reported DGCT 10989/4/0 not found in image management, cannot reallocate image space

Error: 200107023 A Failure had been detected while calling UIR x_ToScanRequest function or srx_sm_generate_action function See Long Description for error variations Tue Oct 19 09:22:20 2004 Error: 200107023 Host: dgct Process: scanRx File: srx_sma_manage_scan.c Line: 688 Function: Data Acquisition : OC Processing srx_sma_prepare_send_request: A Failure had been detected while calling UIRx_ToScanRequest function. Nov 22 08:55:38 2006 Error: 200107023 Host: ct01 Process: scanRx File: srx_sm_event_notify.c Line: 148 Function: Data Acquisition : OC Processing srx_sm_event_notify: A Failure had been detected while calling srx_sm_generate_action function. Tue Aug 14 16:28:25 2007 Error: 200107023 Host: cccgvct1 Process: scanRx File: srx_sma_manage_scan.c Line: 733 Function: Data Acquisition : OC Processing srx_sma_prepare_send_request: A F

Applies to

Models
BrightSpeed
Product
BrightSpeed HPower, Delight, Velocity or MERI HPower HP_P_G4_G_HPT Excel HP_P_G8_G_HPT Edge HP_P_G16_G_HPT Elite HP_P_G16_G_GMV Elite HP_P_P16_G_GMV Elite HP_P_M16_G_HPT BrightSpeed Method of remote software hardware identification check_config swhwinfo List of Supported Console Types : HP_P_G16_ G _HPT Note: G could be GRE, AIO or TIO - check with: cat /usr/g/config/INFO | grep CONSOLETYPE GOC4 AIO TIO List of softwares the product can run: 08BW20.1 07MW11.10 08BW44.1 ( Delight ) 09HW30.4 ( Velocity ) 09BW35.11 ( MERI - Pancake DAS ) FAQ's 1. All BrightSpeed Power Systems use the Performix tube D3186T HP_P_G4_G_HPT (Excel - H-Power Table) HP_P_G8_G_HPT (Edge -H-Power Table ) HP_P_G16_G_HPT (Elite -H-Power Table ) HP_P_G16_G_GMV (Elite -Global Table Mid Value - modified GT1700 no IMS ) Note: G could be GRE, AIO or TIO (check with: cat /usr/g/config/INFO | grep CONSOLETYPE ) BrightSpeed Elite Power Systems will only be sold as Velocity System starting in 2Q/2010. Delight Elite Power being phased out 1Q/2010 BrightSpeed Power 4 & 8 slice systems have been discontinued due to restructuring of the product lineup. Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Software
07MW11.10 for LS 1x2x3x4x5x, HS QX/i, & BSD 4/8/16 07MW11.10 for LS 1x2x3x4x5x, HS QX/i, & BSD 4/8/16 - Original FMI 25389 - Mandatory Safety FMI 25407 - Mandatory Safety - Service Pack 2.0 07MW11.10 for LS 1x2x3x4x5x, HS QX/i, & BSD 4/8/16 This software was updated by Mandatory Safety FMI 25407 ( Service Pack 2.0 for Software Versions 07MW18.4, 07MW11.10, and 07BW08.4), released 6/13/2008 (Initial Release) Always use the most appropriate documentation for your system. The following was taken from DIRECTION 5191771-100, REV 1 and is provided for your reference. Refer to DIRECTION 5191771-100, which can be found on CT LightSpeed CDROM/DVD, or if online at: FCT Service Engineering Tech Pubs Website , or Common Document Library . Software introduction - FIELD MODIFICATION INSTRUCTION / FMI 25389 Class C and Class M - 07MW11.10 Class C Software, p/n 5220446 - 07MW11.10 Class M Software, p/n 5220446-2 NOTE: 07MW11.10 Class C and M CDROMs were distributed to FE’s via a TAB #182 distribution in advance of this FMI. If you have not received them, contact your process leader to ensure you are on TAB #182’s distribution list. Order them through the GEMS Coakleytech website at http://coakleytech.med.ge.com/cgi-win/default.aspx Product Safety Fixes System shutdown due to X/IO error (multiple PSR's) Scan aborts, start scan fails to light, table or cradle not ready due to firmware start up issues on LightSpeed 5.X and BrightSpeed systems (multiple PSR's) Table moving down on it's own Advanced Vessel Analysis gives incorrect measurements if adjust tracking Product Quality Fixes SmartPrep reliability improvements SmartView reliability improvements Start up Shut down reliability improvements Unable to start New Patient (pixMgmt crash) Images fail to Auto View on LightSpeed 4.X 3 IG systems if SmartPrep used Edit Patient fails due to EditPatientData.ct file not updated correctly Alt F4 crashes Display process or Display Manager process Time drift and time jump Multi language Auto voice defaults to reconfig default voice at End Exam Images fail to transfer to a Kodak PACS Images fail to copy or restore from CD/DVD Unable to Save Multi Step protocols in Reformat or Volume Viewer Enhancements / Improvements Auto Transfer by Series for Scouts Auto Transfer for Dose Report Warning message posted that Auto mA is turned “off” when patient orientation and patient entry do not match the scout series Prospective Multi Recon for Cardiac scan type User selectable series description for Retro recon, Reformat and Volume Viewer mA limited if Tube Warm Up skipped on LightSpeed Pro16 systems IQ Snap -- reserve the scan data files related to images with IQ issues for further investigation Full and Partial Anonynous patient levels Scan date added to Large Image Font Screen Save of Exam or Series text page Patient Name and Patient ID added to update parameters in Connect Pro Save Restore System State time optimized
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.