Symptom
Cannot scan with Data Acquisition System Temperature is nearing Upper Limit
Error message
Error: 260002471 Data Acquisition System Temperature is nearing Upper Limit See LD for complete error Thu Feb 16 22:19:58 2006 Host Das Proc: DCB Firmware Error: 260002471 File: ctemptst.c Line: 784 Function: DATA ACQUISTION : Scan Data Acquistion Scan: scan 50024/1/1 Type: Axial Exception: Pri/Soft Time: 22:19:58:760 Log Series: 579 WARNING : Data Acquistion System Temperature is nearing Upper Limit
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 2.X (Plus, QX/i) - Octane Always S-DAS 2.4.2_H2.4M5 LightSpeed 2.X (Plus, QX/i) - Octane (Pegasus or Rhapsode IG) - Always S-DAS Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 2.4.2_H2.4M5 FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 FEATURES/OPTIONS STD/OPTION Cat # Digital Laser Camera I/F B7700PD Smart Helical (s/w only) Standard B7530TK Fast Recon (s/w only) Option B7700JF BMD (requires B7710MP or MN) Option B7710MK Smart Prep Option B7530SP Smart Score ? ? 3D Analysis (Requires D3D) Option B7540TD Navigator (s/w only) Option B7540N DentaScan (s/w only) Option B7540SD Direct 3D (Requires B7710SA) Option B7810SS HIS/RIS Option (h/w and s/w) Option B7500PN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5363182 (replaces 46-220248P3) H1-H2 Plenum Fan FRU GDAS Power Supply Fan FRU and SDAS See long description for details Part #: 5363182 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: H1-H2 Plenum Fan FRU GDAS Power Supply Fan FRU FRU: Yes Status: Orderable Old FRU Compatible Yes Comments Per ECR 2089514, 212008898, 02 Nov 2009. Additional FRU Needed No Old FRU Scrapped No Old FRU Upgradable No Use on Hands No Use Demand History Yes Ultimate Prime 5363182 New FRU Backward Compatible Yes Old FRU Usable Yes Change Type Indirect ECO Number ECR2089514 46-220248P3 FAN: TUBE AXIAL, 115CFM, 115V, 60HZ, Used on H1 and H2 gantries for SDAS Fan / Plenum assembly Part #: 46-220248P3 Part Type: GP CONSUMABLES Part Weight: 1.50 LBR Repairable: No Returnable: No Hazardous Material: N Description: FAN: TUBE AXIAL, 115CFM, 115V, 60HZ FRU: Yes Status: Orderable Comments: FLS - BALL BEARING FAN NECESSARY FOR ROTATION ON CT GANTRY USED ON HELIOS 1 AND HELIOS 2 PER PCN 199761 MS08992
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.