Pular para o conteúdo principal
GE Healthcare · CTLightSpeed2445202601163075100151000

LightSpeed 5.X (60kW)

Symptom

Hardware reset fails with Laser indicator flashing on the Gantry Display board. Hardware reset fails with Laser indicator flashing on the Gantry Display board

Error message

Error: 244520 Scanning hardware has been shut down. NOTE: This is a very generic error. You need to search the log for a more specific error. Fri Apr 21 07:30:53 2006 Error: 244520 Host: oc01 Process: fwmgr File: HwManager.c Line: 1536 Scanning hardware has been shut down. NOTE: This is a very generic error. You need to search the log for a more specific error.

Error: 260116307 Could not send message to 101 or 201 from 101 or 201 (any combination) see long description for error details. Error: 260116307 Could not send message to 101 or 201 from 101 or 201. See examples below: Fri Aug 15 15:01:01 2008 Error: 260116307 Host: tgp Process: tgp_subsys File: lan_watchdog.c Line: 513 Fri Aug 15 15:00:48 2008 Host : Table Gantry Processor Ermes # : 260116307 Exception Class : Secondary Severity : Sec/Soft File : lan_watchdog.c Line# : 513 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Could not send message to 201 from 201. ( TGP from TGP) Event Attempting to send 51001 System Error - 0 Sun Jul 5 13:49:14 2009 Error: 260116307 Host: tgp Process: tgp subsys File: lan watchdog.c Line: 568 Sun Jul 5 14:00:47 2009 Host : Table Gantry Processor Ermes # : 260116307 

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
U8009FF Fuse 3.2 Amp 250v See LD for details Part #: U8009FF Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: FUSE FRU: Yes Status: Orderable
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.