Pular para o conteúdo principal
GE Healthcare · MRDiscovery224860222677982267813226780422678072267801

CWA ID121756. Unable to Scan with Message The IRF3 is not ready.

Symptom

Unable to Scan with Message The IRF3 is not ready. See Long Description Customer is unable to scan due to message in message window located at bottom of Scan Desktop which states “The IRF3 is not ready.->”

Error message

2248602 PDU in Stop Mode. Emergency Stop button was pressed or PDU 24 Volt supply has failed. 2248602 PDU is in Stop mode. Emergency Stop button was pressed or PDU 24 Volt supply has failed.

2267798 Safe mode startup of the XGD (XGA-SB X axis) failed. Check EMO reset and perform a TPS reset. See Long Description 2267798 "Safe mode startup of the XGD (XGA-SB X axis) failed. Check EMO reset and perform a TPS reset. If this error occurs again, contact your service representative.-->"

2267813 Safe mode startup of the XGD (XPS-SB Z axis) failed. Check EMO reset and perform a TPS reset. See Long Description 2267813 "Safe mode startup of the XGD (XPS-SB Z axis) failed.\n\ Check EMO reset and perform a TPS reset. If this error occurs again, contact your service representative.->"

2267804 Safe mode startup of the XGD (XGA-SB Z axis) failed. Check EMO reset and perform a TPS reset. See Long Description 2267804 "Safe mode startup of the XGD (XGA-SB Z axis) failed. Check EMO reset and perform a TPS reset. If this error occurs again, contact your service representative.->"

The IRF3 is not ready See Long Description "The IRF3 is not ready" This message appears in the customer message window 

Applies to

Models
Discovery
Product
Discovery MR750. Also called DVMR. This product is compatible with 3.0T 20.0, 20.1_IB1, IB2, 21.0 and DV22.0 software only
Software
20.0 Also called DVMR. This software is compatible with 3.0T Signa MR750 systems only
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.