LightSpeed 5.X (60kW)

Symptom

Gantry tilts slowly 1 degree and stops - table moves very slowly 5mm and stops. Gantry tilts slowly 1 degree and stops - table moves very slowly 5mm and stops

Error message

Error: 182427 Gantry Back / Forward Touch Sensor Error See Long Description for complete Error Error: 182427 Gantry Tilt Back or Gantry Tilt Forward Sensor Error Error: 182427 Host: ETC Process: Elevation File: emcTouchSensor.c Line: 125 Function: PATIENT ANATOMY POSITIONING : Elevation - Tilt Scan: 12033/1/0 Type: None/Unknown Exception Level: Pri/Soft Time: 14:58:23:525 Log Series: 344 The Gantry BACK Touch Sensor Error : Console Pushbutton Remote Tilt is DISABLED. Use Gantry Controls to Tilt the Gantry. Current Position 14660 (.01 mm). Elev Dir: 1 (0 = Up 1 = Down) (Elev Cmd: 1 TouchSensor: 0x8000). Same Error for the Forward Touch sensor

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2277405 Front / Small Gantry Tilt Forward Sensor / Interference Switch used on H2 gantry See Long Description for Details 2277405 Front / Small Gantry Tilt FORWARD Sensor / Interference Switch used on H2 Part #: 2277405 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: PATIENT SENSOR, SMALL FRU: Yes Status: Orderable Comments : FLS - SAFETY TILT INTERFERENCE SWITCH FRONT USED ON H2 COMMON TO MFG. MS26561 DSM 1/7/01 COLLECTION OF DEFECTIVES START7/23/01 MS5277 FLS - PLEASE CONVERT SPH ORDER TO OVERNIGHT SHIPMENT IF PART AVAILABLE IN MEMPHIS AND IF FEDEX SERVICE (INCLUDING 8:00 AM SERVICE) MEETS FE NEED TIME CRITERIA
Area
CT

Full cause & step-by-step resolution