GE Healthcare · CTLightSpeed260118168186218

Scanner Hardware Stopped Scan

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 260118168 (40-0105) Rotation Phases unbalanced See LD for error explanation Code 40 Error code: 40-0105H Message/explanation: Rotation Phases unbalanced The amplitude difference of the current between main and auxiliary is too large. Potential cause: 1. Tube stator winding problem 2. Incorrect wiring (one wire open) 3. Phase shift capacitors inverted or wrong value or not connected 4. Rotation board Recommended action/ Troubleshooting guide: Disconnect rotation-to-tube-stator cable and check stator windings impedance. If not OK, replace tube. Check wiring from rotation board to tube. Check that phase shift capacitors are correctly connected Check that phase shift capacitors are not in short-circuit Run Rotation diagnostic (See diagnostic section). If test fails, replace Rotation board Tue May 9 09:26:56 2006 Error: 186218 Host: IRIS Process: tgp_out File: tgp_in_input.c Line: 335 XG Error phase is 3 code is 40. detail error code : 40 - 0105H Rotation error : Rotation Phases unbalanced

Applies to

Models
LightSpeed
Product
LightSpeed RT16 & RT16 Xtra and Vision WB_H_G16_G_GTL = RT16, Xtra WB_H_G16_G_GTL650 = Vision WB_H_G16_G_GTM = RT16, Xtra or Vision LightSpeed RT16 & RT16 Xtra with GRE and GDAS16 gantry Links to the Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2324118 JH4 AUXILIARY EQUIPED V2 - Used on HP and Brightspeed gantries. Not used on VCT See long description for details Part #: 2324118 Part Type: GP CONSUMABLES Part Weight: -98999.01 LBR Repairable: No Returnable: No Hazardous Material: N Description: JH4 AUXILIARY EQUIPED V2 FRU: Yes Status: Orderable Comments: FLS - JH4 AUXILIARY EQUIPED V2. TO BE USED PER HPOWER PROGRAM RSPL. MS20544 ON 2/12/03.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.