GE Healthcare · CTLightSpeed18221360046
LightSpeed 4.X - Octane
Symptom
Cradle will not drive in or out. Cradle will not drive in or out
Error message
Error: 182213 Velocity tracking error See Long Description for the full error code Tue Mar 15 15:28:05 2005 Error: 182213 Host: ETC Process: Cradle File: CmdMotion.c 1.33 7/23/96 Line: 92 Function: PATIENT ANATOMY POSITIONING : Cradle Drive Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 15:28:05:770 Log Series: 37 HW error: Velocity tracking error. Expected Velocity 500 (.001 mm/sec) Actual Velocity 9382 (.01 mm/sec) Current Position 60046 (.01 mm) Velocity Tolerance 250 (.001 mm/sec) Cradle Dir: 1 (0 = Out 1 = In [Gantry]).
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 4.X - Octane sw_version: 405I.2_H4.0M5 LightSpeed 4.X - Octane Gantry Model # 2339985 Console Model #'s 2341923 and 2341923-2(D3D) check_config shows LightSpeed 16 / LightSpeed 16 Product: LightSpeed16 sw_version: 405I.2_H4.0M5 or newer machine_type: Octane host_id: 3232297858 tube_type: 15 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required