GE Healthcare · CTLightSpeed18220870636654301049466531
LightSpeed 4.X - GOC2
Symptom
Cradle will move a few inches and stop. Cradle will move in a few inches and stop
Error message
Error: 182208 Alignment error between Encoder and Potentiometer. Potentioment Position 70636 (.01 mm) Encoder Position 65430 (.01 mm) Tolerance 2540 (.01 mm) See the LD for the complete error code Error: 182208 Alignment error between Encoder and Potentiometer. Tue Aug 5 15:05:36 1997 Suite: PCT2 Host: ETC Proc:Cradle Error: 182208 File: cmcEncoderRef.c 1.8 8/27/ Method: No Method Line: 152 Function: PATIENT ANATOMY POSITIONING : Cradle Drive Scan: 10494/1/0 Type: None/Unknown Exception Level: Pri/Soft Time: 15:05:35:935 Log Series: 27 HW error: Encoder C Pulse Alignment error with current Potentiometer reading. Current Position 66531 (.01 mm). Encoder Alignment Minimum (1456) and Maximum (1658) Potientiometer Window based on Last Potentiometer Reading (1658).
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 4.X - GOC2 405L.8_H4.2M4 LightSpeed 4.X with GOC2 and MDAS 16 Gantry H4 Gantry Model # 2339985 and Global Operator Console 2 Model #'s 2366438 and 2366438-2(Hino) with HP8000 CPU (or newer) and Pegasus Recon ICE Box Product: LightSpeed16 405L.8_H4.2M4 CPU: GenuineIntel Intel(R) Xeon(TM) CPU 2.66GHz 4 2657 MHz i686 Processors Main memory size: 2048 Mbytes [root@presbyct2]# sg_map -i /dev/sg0 /dev/sda SEAGATE ST336753LW HPS2 /dev/sg1 /dev/sdb SEAGATE ST373453LW HPS1 /dev/sg2 /dev/sdc SONY SMO-F551-SD 1.05 /dev/sg3 /dev/scd0 MATSHITA DVD-RAM SW-9574S AZX8 /dev/sg4 /dev/scd1 HL-DT-ST DVD-ROM GDR8163B 0B15 [root@presbyct2]# df -h Filesystem Size Used Avail Use% Mounted on /dev/sda2 4.0G 1.7G 2.3G 41% / /dev/sda1 27M 12M 14M 44% /boot none 1008M 0 1007M 0% /dev/shm /dev/sda6 14G 3.5G 10G 25% /usr/g /dev/sda7 9.8G 753M 9.0G 8% /usr/g/sdc_image_pool /dev/sdb1 68G 28G 40G 41% /usr/g/sdc_image_pool2 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2269175-3 LONGITUDINAL ENCODER ASM (Used on HSA, CTi, H1 QXi, Plus, Ultra, 16 slice H4 and HiSpeed QXi.) replaces 2269175-2. See long description for details Part #: 2269175-3 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Longitudinal Encoder Assembly with Dynapar encoder FRU: Yes Status: Orderable Comments: Per ECR 2036924, 212008898, 12/4/2006. ** FE Assist ** If no stock or unable to order the 2269175-3 , order 2269175-2. (212001261 5/22/07) Part #: 2269175-2 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Longitudinal Encoder Asm. NOTE:This asm. has LCC encoder/tested and sensor gap is 0.06-0.08mm FRU: Yes Status: Orderable Comments: Per ECR 2036924, 212008898, 12/4/2006
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.