LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

Pop up during fastcal indicates potential image quality issue. See Long Description for error stream example. Pop up error posted to operator during fastcal indicated potential image quality issue. Errors 230007000 230007002 230007014 posted in gesyslog. Exclusion test, open channel failed Z-axis exclusion and proximity criteria. Fri Oct 28 09:32:03 2005 Error: 230007000 Host: presbyplano Process: badchanmgr File: BadChannelMap.cxx Line: 616 Exclusion Test - Row 64:32B, Channel 695, Module 44 Open Channel* FAILED Z- AXIS EXCLUSION CRITERIA *. Fri Oct 28 09:32:03 2005 Error: 230007002 Host: presbyplano Process: badchanmgr File: BadChannelMap.cxx Line: 638 Exclusion Test - Row 64:32B, Channel 695, Module 44 Open Channel* FAILED PR OXIMITY CRITERIA *. Fri Oct 28 09:32:03 2005 Error: 230007014 Host: presbyplano Process: badchanmgr File: badchanmgr.cxx Line: 1194 Uncorrectable detector channels have been found for scan 50029/2/2

Error message

Error: 230007000 Exclusion Test - Row 48:16B, Channel 552, Module 35 Open Channel* FAILED Z-AXIS EXCLUSION See LD for complete error. Tue Jul 19 03:54:27 2005 Error: 230007000 Host: ct01 Process: badchanmgr File: BadChannelMap.cxx Line: 533 Exclusion Test - Row 48:16B, Channel 552, Module 35 Open Channel* FAILED Z-AXIS EXCLUSION CRITERIA *.

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Component
5136400-3 FRDM Center Module Kit for VDAS. see long description. Part #: 5136400-3 VCT CENTER DIGITAL MODULE KIT for VDAS Part #: 5136400-3 Part Type: GP REPAIRABLES Part Weight: 0.00 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: VDAS CENTER DIGITAL MODULE KIT for VCT FRU: Yes Status: Orderable ** FE ASSIST ** This is single center slot (27,28,29,30,31) VDAS64 Detector Module for field replacement. Refer to Service Note SN05091210e or later revision for any special instructions
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.