Affter Infinia Hawkeye 1 upgrade to Hawkeye 4, X-Ray scan fails to start, but diagnstics pass

Symptom

Affter Infinia Hawkeye 1 upgrade to Hawkeye 4, X-Ray scan fails to start, but diagnstics pass Affter Infinia Hawkeye 1 upgrade to Hawkeye 4, X-Ray scan fails to start, but diagnstics pass. After upgrade Infinia VC Hawkeye 1 to Infinia VC Hawkeye 4, found all diagnstic for Xray TTH Generator and Detector, DAS, all passed successfully, but X-ray scan failed, checked error log, found error regarding no offset data.

Error message

CT_ERROR_AIA_OFFSET,6105,ERROR:.: Error AIA could not stabilize offsets.:: Cannot prepare data needed for Xray image correction

Applies to

Models
Infinia
Product
Infinia HE4 Infinia II Hawkeye 4 & Infinia II GP (Infinia running Windows XP software)
Component
*X-Ray (Generator, DAS, Controller) *X-Ray (Generator, DAS, Controller)
Area
Nuclear Medicine

Full cause & step-by-step resolution