GE Healthcare · CTLightSpeed1833401831825000144858
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 183340 The rotor control board has detected HVDC Bus voltage when none expected. See the Long Description for the complete error code Tue Jul 1 11:53:31 2003 Error: 183340 Host: OBC Process: Generator File: gen_power.c 1.13 Line: 337 Function: SYSTEM MONITORING : HVDC Monitor Scan: 858/5/1 Type: Helical Exception Level: Pri/Soft Time: 11:53:31:975 Log Series: 675 The rotor control board has detected HVDC Bus voltage when none expected. The Gentry I/O indicates the backup pilot relay is de-energized. HVDC Bus voltage: 46V Expected: 0V Limit: 24V Error: 183182 I/O board pilot relays are closed. PDU pilot relay not energized. HVDC Rail Voltage is xxxx See Long Description for more details. Error: 183182 I/O Bd. pilot relays are closed. PDU pilot relay not energized. HVDC is xxxx. ** Note the HVDC Rail Voltage. It is important in troubleshooting Tue Sep 21 06:13:59 2004 Error: 183182 Host: OBC Process: Generator Function: SYSTEM POWER CONTROL: HVDC Control Scan: 50001/1/1 Type: Axial Exception Level: Pri/Most Severe I/O board pilot relays are closed. PDU pilot relay not energized BCSTAT = 0 HVDC Rail voltage = 0V Sun Dec 12 18:11:57 2004 Error: 183182 Host: OBC Process: Generat
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2139289-2 CPDU Control Bd. Replaces part 2139289 Used In CPDU - 2269902. 2139289-2 CPDU Control Bd. for Compact PDU Part #: 2139289-2 Part Type: GP REPAIRABLES Part Weight: 5.35 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: 2139289-2 : PDU CONTROL BOARD W/ REW FRU: Yes Status: Orderable Comments: FLS - PDU CONTROL BOARD REWORK ASSEMBLY. THIS P/N 2139289-2, IS P/N T3202KF (2139289) WITH A REWORKED SUB-ASSEMBLY. ADDED PER PCN 220525 BY MS20544 ON 7/17/02. 9/10/02- 2139289-2 REPLACES 2139289 PER PCN 220525. MS12464 11/08/04 SSO# 212001261 FE's are to return defectives and unused/consigned back via FED Ex over night Item #16 A6 panel is the control board
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required