GE Healthcare · CTLightSpeed183318
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane
Symptom
Warning message posted in the gesyslog with no customer complaints of scan aborts Warning messages posted in the gesyslog with no customer complaints of scan aborts
Error message
Error: 183318 The HEMRC output current is out of 35.0 percent WARNING tolerance. See Long Description for complete error. Error: 183318 The HEMRC output current is out of 35.0 percent WARNING tolerance. Wed Jun 11 13:53:05 2008 Host: OBC Proc: Generator Error: 183318 File: hemrc_manager.c 1.14 Line: 1108 Function: XRAY GENERATION : Tube Rotor Control Scan: 6/1/1 Type: Scout Exception Level: Pri/Soft Time: 13:53:05:670 Log Series: 78 The HEMRC output current is out of 35.0 percent WARNING tolerance. Mode = Normal Monitored state = 0 (0=Accel 1=Trans 2=Run state) Expected current = 11.999 A Actual current = 7.750 A. Note: This error is a warning message not an abort able error: if Expected current is greater than Actual current during Acceleration state this may not be a problem (Tube bearings better than expected) if Expected current is less than Actual current this may be a problem in the HEMRC circuit
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required