GE Healthcare · CTHiSpeed183318

HiSpeed CT/i 4.X with SGI Indigo Console, HiSpeed Advantage Gantry

Symptom

Warning message posted in the gesyslog with no customer complaints of scan aborts Warning messages posted in the gesyslog with no customer complaints of scan aborts

Error message

Error: 183318 The HEMRC output current is out of 35.0 percent WARNING tolerance. See Long Description for complete error. Error: 183318 The HEMRC output current is out of 35.0 percent WARNING tolerance. Wed Jun 11 13:53:05 2008 Host: OBC Proc: Generator Error: 183318 File: hemrc_manager.c 1.14 Line: 1108 Function: XRAY GENERATION : Tube Rotor Control Scan: 6/1/1 Type: Scout Exception Level: Pri/Soft Time: 13:53:05:670 Log Series: 78 The HEMRC output current is out of 35.0 percent WARNING tolerance. Mode = Normal Monitored state = 0 (0=Accel 1=Trans 2=Run state) Expected current = 11.999 A Actual current = 7.750 A. Note: This error is a warning message not an abort able error: if Expected current is greater than Actual current during Acceleration state this may not be a problem (Tube bearings better than expected) if Expected current is less than Actual current this may be a problem in the HEMRC circuit

Applies to

Models
HiSpeed
Product
HiSpeed CT/i 4.X HiSpeed CT/i 4.x with Zeus HSA Gantry and SGI Indigo Console HiSpeed CT/i with:HSA GANTRY CT/I (RHAPSODE) CONSOLE (Indigo Computer) Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.