GE Healthcare · CTLightSpeed26011631650119244520
LightSpeed 5.X (60kW)
Symptom
Hardware reset fails with Laser indicator flashing on the Gantry Display board. Hardware reset fails with Laser indicator flashing on the Gantry Display board
Error message
Error: 260116316 Communications are DOWN between subsystems See LD for full error Mon Jun 13 14:34:09 2005 Error: 260116316 Host: tgp Process: tgp_subsys File: lan_watchdog.c Line: 904 Mon Jun 13 14:34:03 2005 Host : Table Gantry Processor Ermes # : 260116316 Exception Class : Safe State Severity : Primary File : lan_watchdog.c Line# : 904 Function : No System Function Reported Scan Type : None/Unknown Scan: 7754/5/0 Communications are DOWN between subsystems Or Sun Dec 17 08:26:45 2006 Error: 260116316 Host: tgp Process: tgp_subsys File: lan_watchdog.c Line: 996 Sun Dec 17 08:26:39 2006 Host : Table Gantry Processor Ermes # : 260116316 Exception Class : Safe State Severity : Secondary File : lan_watchdog.c Line# : 996 Function : No System Function Reported Scan Type : None/Unknown Scan: 50119/1/0 Slip-Ring Communications are DOWN between subsystems Error: 244520 Scanning hardware has been shut down. NOTE: This is a very generic error. You need to search the log for a more specific error. Fri Apr 21 07:30:53 2006 Error: 244520 Host: oc01 Process: fwmgr File: HwManager.c Line: 1536 Scanning hardware has been shut down. NOTE: This is a very generic error. You need to search the log
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5133765 GDAS Power Supply Upgrade Kit: to Astec Power Supplies See the LD for FEMC part comments 5133765 GDAS Power Supply Upgrade Kit Part #: 5133765 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: GDAS Power Supply Upgrade Kit: to Astec Power Supplies FRU: Yes Status: Orderable This part is Service replacement Kit for GDAS power supplies 2330352-4 and 2330352-3. Single kit will replace BOTH analog and digital power supplies at same time. ms13452 050420 Note: (1/2/06 SMF) Comments from Dennis L.Badgerow, The Kit is shipped in a box the size of an x-ray tube, only weighs 110 lbs but is really big should ship directly to the site if possible. Next, the directions say one FE for two hours, it took two FE's nearly five hours with no interruptions.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.