Pular para o conteúdo principal
GE Healthcare · CTLightSpeed2300000652300000675000416800015000

LightSpeed 5.X (60kW)

Symptom

Scanning aborted with Buffer overrun detected during scan data transfer

Error message

Error: 230000065 Buffer overrun detected during scan data transfer. See long description for full error Error: 230000065 Buffer overrun detected during scan data transfer. This is generally not a hardware issue and will not be remedied by replacing or otherwise servicing the gantry, fiber-optic cable or DIP card. Interrupt bit: 6 Violation count: 1 Scan aborted.

Error: 230000067 Unexpected event occurred during scan data transfer See the long description for the full error code Sat Apr 17 20:34:13 2004 Error: 230000067 Host: darc Process: dataacq File: dip_thread.cxx Line: 390 Running diagnostic: Scan Data Path Functional Function: Data Acquisition : Scan Data Acquisition Scan Type: Xray-on Static Scan: 50004/4/1 Scan Seq Id: 168000 Unexpected event occurred during scan data transfer. Event bit: 7 Count: 1 Description: Global bus error: read attempted from the Test Data FIFO Scan aborted. OR Tue Mar 6 07:13:50 2007 Error: 230000067 Host: darc Process: dataacq File: vdip.cxx Line: 817 Function: Data Acquisition : Scan Data Acquisition Scan Type: Cine Scan: 50004/6/1 Scan Seq Id: 15000 Unexpected event occurred during scan data transfer. Event bit: 22 Count: 1 Description: DMA inter

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues