GE Healthcare · CTLightSpeedBrightSpeed26011821410132
LightSpeed 5.X (60kW)
Symptom
CTx00188 Proactive svc call by PDS This Issue was detected by the Digital Services (DS) rule engine. This Issue is only applicable when detected by the Rules Engine and should not be reused in Knowlege objects for normal reactive service. CTx00188 Proactive svc call by PDS This Issue was detected by the Digital Services (DS) rule engine. This Issue is only applicable when detected by the Rules Engine and should not be reused in Knowlege objects for normal reactive service.
Error message
Error: 260118214 (60-0317) Spit Ratio error See Long Description for complete error. Fri Dec 16 23:26:55 2005 Error: 260118214 Host: orp Process: orp_subsys File: gen_exec.c Line: 2150 Fri Dec 16 23:26:54 2005 Host : Jedi Ermes # : 260118214 Exception Class : Abort Severity : Primary File : gen_exec.c Line# : 2150 Function : No System Function Reported Scan Type : Axial Scan: 10132/2/1 (60-0317)Spit Ratio error. PPC board has detected the rate of tube spits during exposure has reached the limit (see theory of operation, software section).^C Z6,Z7= 4h,0h
Applies to
- Models
- LightSpeed, BrightSpeed
- Product
- LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5126991 INVERTER MID PWR NP V3 - used with performix tube ( replaces part number 2242442 ) See long description for details Part #: 5126991 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: INVERTER MID POWER ASSY V3_FRU FRU: Yes Status: Orderable Comments: Not Available Part #. 2242442 ASSY INVERTER MID PWR NP V3 Part Type: GP CONSUMABLES Part Weight: 54.67 LBR Repairable: No Returnable: No Hazardous Material: N Description: ASSY INVERTER MID PWR NP V3 FRU: Yes Status: Orderable Comments: FLS - ASSY INVERTER MID PWR NP V3 U/O JEDI NPV3 (YMS ZXI) MS08992 8/24/99 M3 FW 44 '99 DO NOT DISCOUNT - PER SHAKUN ** FE ASSIST ** this inverter is used on Jedi V3 Jedi for Performix Tube (212001261 7/30/08)
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required