Pular para o conteúdo principal
GE Healthcare · CTLightSpeed26011818550103

LightSpeed 5.X (60kW)

Symptom

CTx00188 Proactive svc call by PDS This Issue was detected by the Digital Services (DS) rule engine. This Issue is only applicable when detected by the Rules Engine and should not be reused in Knowlege objects for normal reactive service. CTx00188 Proactive svc call by PDS This Issue was detected by the Digital Services (DS) rule engine. This Issue is only applicable when detected by the Rules Engine and should not be reused in Knowlege objects for normal reactive service.

Error message

Error: 260118185 (50-0204) Filament permanent open circuit See Long Description for complete error. Host: orp Process: orp_subsys File: gen_exec.c Line: 2376 Wed Mar 21 08:25:44 2007 Host : Jedi Ermes # : 260118185 Exception Class : Abort Severity : Primary File : gen_exec.c Line# : 2376 Function : No System Function Reported Scan Type : Axial Scan: 50103/8/6 (50-0204)Filament permanent open circuit. Issued by the heater board when an open has been detected and 3 restarts have been tried without success within 100 ms. ^C Z6,Z7= 97h,0h

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2361419 Programmed Heater Board (eqvt 2409049) inside Jedi Auxiliary Module 2324118 See LD for Part Details - please read the note at the bottom. Part number for Heater (Filament) Board below: Part #: 2361419 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: PROGRAMMED HEATER 6P04 FRU: Yes Status: Orderable Part #: 2409049 Part Type: GP CONSUMABLES Part Weight: 0.00 KGM Repairable: No Returnable: No Hazardous Material: N Description: FRU-PROGRAMMED HEATER V6.04 FRU: Yes Status: Orderable Parts Offered as: Current: 2409049 Harvested: Not Available Exchange: Not Available Comments: added per to ECR2088514 by sso#305010201 in Jan 15, 2010 Note: * Order 2409049 to replace 2361419.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.