Pular para o conteúdo principal
GE Healthcare · CTLightSpeed185122182902

LightSpeed 4.X - GRE

Symptom

Table will not move. Table will not move

Error message

Error: 185122 24V Power Supply Out Of Tolerance Sun Sep 5 16:04:34 2004 Error: 185122 Suite: RP21 Host: ETC Process: Control File: etcPowSupTest.c 1.4 2/16/99 Method: No Method Line:117 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 3588 Log Series: 2 HW error: 24V Power Supply Out Of Tolerance: A/D reading = 0x4 Calc Voltage = 0.03 (Volts) Tolerance = 3 (Percentage) INFO: If the supply is low the problem may be that the Xray Drive Power is OFF

Error: 182902 Emergency Off button has been pressed Sun Sep 5 16:04:39 2004 Error: 182902 Suite: RP21 Host: ETC Process: Table Controls File: tc emergency.c 1.6 2/16/99 Method: No Method line: 89 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 4016 Log Series: 22 Emergency Off button has been pressed

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2372371 Left Table Top Cover - used on LS 2.x - 5.x tables See LD for details 2372371 HPower Left Table Top Cover 2372371 1.000 Yes COVER TABLE TOP LEFT left...... Part #: 2372371 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: COVER TABLE TOP LEFT FRU: Yes Status: Orderable Comments: FLS - TABLE COVER, TOP LEFT. ADDED PER CT PCN 220864. THIS PART REPLACES P/N 2236720 ONLY ON HPOWER SYSTEMS . KEEP USING OLD P/N FOR ALL OTHER PRODUC TS. MS20544 3/31/03. Additional Left Table Top Cover information for 1.x, 2.x, 3.x, 4.x and HiSpeed QXi Systems. Part #: 2236720 <--Use this part # for 1.x, 2.x, 3.x, 4.x and HiSpeed QXi Systems Part Type: GP CONSUMABLES Part Weight: 16.01 LBR Repairable: No Returnable: No Hazardous Material: N Description: COVER ASSEMBLY TABLE TOP - LEFT FRU: Yes Status: Orderable Comments: FLS - CAN ALSO ORDER OLD COVER 2149735-2 GPC TABLE MOVE TO CHINA 5/24/0NOW ASIA SOURCED. PER PCN 220864, THIS PART HAS BEEN REPLACED BY P/N 2372371, IN ORDER TO BE USED ON HPOWER SYSTEMS. ALSO SOURCED BY ASIA. MS205444 3/31/03. DO NOT DISCOUNT - PER SHAKUN RIGHT: 2236720-2 ONE USE LIGHTSPEED FLS - PLEASE CONVERT SPH ORDER TO OVERNIGHT SHIPMENT IF PART AVAILABLE IN MEMPHIS AND IF FEDEX SERVICE (INCLUDING 8:00 AM SERVICE) MEETS FE NEED TIME CRITERIA 2372372 Right Table Top Cover - used on LS 2.x - 5.x tables See LD for details 2372372 HPower Right Table Top Cover Part #: 2372372 Part Type: GP CONSUMABLES Part Weight: 72.75 LBR Repairable: No Returnable: No Hazardous Material: N Description: COVER TABLE TOP RIGHT FRU: Yes Status: Orderable Comments: FLS - COVER TABLE, TOP RIGHT. REPLACES P/N 2236720-2 ONLY ON HPOWER SYSTEMS, PER CT PCN 220864. MS20544 3/31/03. Addtional Right Table Top Cover Information: Part #: 2236720-2 <-- Use this part # for 1.x, 2.x, 3.x, 4.x and HiSpeed QXi Systems Part Type: GP CONSUMABLES Part Weight: 25.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: COVER ASSEMBLY TABLE TOP - RIGHT FRU: Yes Status: Orderable Comments: FLS - COVER ASSEMBLY TABLE TOP-RIGHTU/O LIGHTSPEED &amp; PET TABLE 6/30/99 MS08992 THIS REPLACED THE 2149735 PER CHRIS DOERING MSA1248 10/99 GPC TABLE MOVED TO CHINA 5/24/01 NOW ASIA SOURCED THIS P/N HAS BEEN REPLACED BY P/N 2372372 (ONLY APPLICABLE ON HPOWER SYSTEMS) PER PCN 220864. MS20544 3/31/03. DIMS: 90&quot; LONG; 70&quot; WIDE FLS - PLEASE CONVERT SPH ORDER TO OVERNIGHT SHIPMENT IF PART AVAILABLE IN MEMPHIS AND IF FEDEX SERVICE (INCLUDING 8:00 AM SERVICE) MEETS FE NEED TIME CRITERIA
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.