Pular para o conteúdo principal
GE Healthcare · CTHiSpeed

Sytec Synergy Series

Symptom

Center Smudge Artifact. Goes away with Fast Cals but returns same day or next day. See Long Description for artifact. Center Smudge Artifact, can be light or dark, usually a horseshoe shape or C shape, sometimes appears as a complete circle band. Reason for change in appearance is as the tube radiates heat to the Collimator Filter it expands changing the size of the crack and then the current calibrations can no longer correct for the crack.

Applies to

Models
HiSpeed
Product
Sytec Synergy Series - Synergy S, Synergy, Synergy Plus, Sytec SRi/Legato XMj CT Sytec Synergy Series - Synergy S, Synergy, Synergy Plus, Sytec SRi/Legato XMj CT Sytec Synergy Series - Synergy S, Synergy, Synergy Plus, Sytec SRi, Legato XMj For Synergy S to Synergy Upgrade Kit (2185628) OC Label (2108440) GANTRY Label (2108443). For Synergy to Synergy Plus Upgrade Kit (2185630) GEMS–E: OC Label (2108440) GANTRY Label (2108443) Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
P9144NF COLLIMATOR FILTER used on YMS Synergy P9144NF COLLIMATOR FILTER P9144NF COLLIMATOR FILTER Part #: P9144NF Part Type: GP CONSUMABLES Part Weight: 0.91 KGM Repairable: No Returnable: No Hazardous Material: N Description: FILTER FRU: Yes Status: Orderable
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.