GE Healthcare · CTLightSpeed2600031212600031002600010235072326000235850011
LightSpeed 2.X (Plus, QX/i) - Octane
Symptom
"Scan hardware Stopped Scan" with collimator errors. "Scan hardware Stopped Scan" with collimator errors
Error message
Error: 260003121 The Aperture failed to move to the requested position. See LD for complete error Mon Dec 6 12:55:55 2004 Error: 260003121 Host: HSC Process: CCB Firmware File: aperture_move.c Line: 225 Running diagnostic: Collimator Functional Function: XRAY GENERATION : Xray Collimation and Filtration Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 12:55:56:165 Log Series: 63 The Aperture failed to move to the requested position. Error: 260003100 CAM B Motion With Out Command Exceeded The Error Tolerance. Desired Position : 251um 797cnt Current Position : 415um 763cnt Error Tolerance: 44um Mon Dec 6 13:01:26 2004 Error: 260003100 Host: HSC Process: CCB Firmware File: ccb_drive_timers.c Line: 277 Running diagnostic: Collimator Functional Function: XRAY GENERATION : Xray Collimation and Filtration Scan: 0/0/0 Type: None/Unknown Exception Level: Sec/Soft Time: 13:01:27:185 Log Series: 42 CAM B Motion With Out Command Exceeded The Error Tolerance. Desired Position : 251um 797cnt Current Position : 415um 763cnt Error Tolerance: 44um Do Not Use - use one of the other two Error: 260001023 See Long Description for complete error NOTE: The error message code included in t
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 2.X (Plus, QX/i) - Octane Always S-DAS 2.4.2_H2.4M5 LightSpeed 2.X (Plus, QX/i) - Octane (Pegasus or Rhapsode IG) - Always S-DAS Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 2.4.2_H2.4M5 FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 FEATURES/OPTIONS STD/OPTION Cat # Digital Laser Camera I/F B7700PD Smart Helical (s/w only) Standard B7530TK Fast Recon (s/w only) Option B7700JF BMD (requires B7710MP or MN) Option B7710MK Smart Prep Option B7530SP Smart Score ? ? 3D Analysis (Requires D3D) Option B7540TD Navigator (s/w only) Option B7540N DentaScan (s/w only) Option B7540SD Direct 3D (Requires B7710SA) Option B7810SS HIS/RIS Option (h/w and s/w) Option B7500PN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 46-296259P6 Rotating Coupling See Long Description for FEMC and pics 46-296259P6 ROCOM COUPLING_BORES 5 MM pictures Part #: 46-296259P6 Part Type: GP CONSUMABLES Part Weight: 0.10 LBR Repairable: No Returnable: No Hazardous Material: N Description: ROCOM COUPLING_BORES 5 MM, 0.250 IN. FRU: Yes Status: Orderable Comments: FLS - ADAPTER ADDED PER PCN #220365.USED ON GANTRY. PER PCN COMMENTS, FRU WILL BE CHANGED FROM N TO 1. MSA8429 01/15/02 COUPLING OF 2 SHAFTS FOR COLLIMATOR MOTOR MS10147 01/02 FOR ROTATING APTERETURE
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.