Pular para o conteúdo principal
GE Healthcare · CTLightSpeed23001301524504730000230005024230006530000000

LightSpeed 4.X - Octane

Symptom

Multiple recon retries and suspended images during an exam. See long description for details Multiple recon retries and suspended images during an exam. To get the images to recon you would have to resume the queue or reboot the system for the images to recon.

Error message

Error: 230013015 xx DAS views of 100 retrieved from the scan database failed checksums See Long Description for great info on this error Problem: The Following errors all point to a scsi Scan data disk issue. Using the gesys_Ct01.log and iceConsole.log is a must, together both logs point to the Scan data disk chain. Check DIP stats if stats are clean the das data is good. The iceConsole log is a little difficult to read and hard to correlate to specific dates and times. --------------gesys_CT01.log--------------- Sun Jan 13 20:33:00 2002 Error: 245047 Host: gect88 Process: scanmgr File: scan_sequence.cxx Line: 1377 Function: System Control : Scan Control Scan Type: Helical Scan: 0/0/0 Scan Seq Id: 30000 Unable to allocate scan files. Sun Jan 13 20:33:05 2002 Error: 230013015 Host: ice Process: rebin File: RebinMDASData.cxx Line: 332 Function: Image Generation : Data Restore Job Type: Prospective Helical Scan: 31/2/1 Image: 31/2/56 Recon Seq Id: 173 xx DAS views of 100 retrieved from the scan database failed checksums <-- DAS Das fails checksum from Scan data disk!!!! Sun Jan 13 20:33:05 2002 Error: 230005024 Host: ice-sm-gate Process: sdcctrl File: sdcctrl_events.cxx Line: 358 imag

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - Octane sw_version: 405I.2_H4.0M5 LightSpeed 4.X - Octane Gantry Model # 2339985 Console Model #'s 2341923 and 2341923-2(D3D) check_config shows LightSpeed 16 / LightSpeed 16 Product: LightSpeed16 sw_version: 405I.2_H4.0M5 or newer machine_type: Octane host_id: 3232297858 tube_type: 15 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Software
405I.2_H4.0M5 for LightSpeed 4.X (LightSpeed 16) with H16 Console (GOC1)(2341923, 2341923-2) - showprods returns 405I.2_H4.0M5 for LightSpeed 16 with GOC1 SGI Octane CPU 405I.2_H4.0M5 for LightSpeed 16 with GOC1 SGI Octane CPU Class C Class M Application Software Set CD Part No. Ver. CD Label CD Part No. Ver CD Label Set Part No. Apps. CD Part No Apps. CD Label showprods/ Service Desktop Intro. Date Comment 2395172 1104I.2_ H4.0M5 CLASSC LightSpeed MDAS 4/8/16/ HiSpeed QXi Class C Software 2395173 1204I.2_ H4.0M5 CLASSM LightSpeed/ HiSpeed QX/I Class M Software 2395176 2395170, 2395171, 2399867 LightSpeed 16 Applications Software (Disk 1, Disk 2 and Patch A) 405I.2_H4.0M5 FW47-03 LS2002M5. IRIX only. Class C/M Common w/ LS 2X, 3X, & 4X.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required