GE Healthcare · CTLightSpeedVCT290010005290009012290009013
LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
A Pink Pop up message is reported on boot-up that Recon times may be slower - scanning may continue See Long description to see the Pink Pop up message A Pink Pop up message is reported on boot-up that Recon times may be slower
Error message
Error: 290010005 Socket error: socket is terminated because IG dies (error text varies) See Long Description for error examples. Tue Mar 16 10:40:09 2004 290010005 darc data_restore request_object.cxx 503 Socket error: socket is terminated because IG dies Fri Nov 16 19:22:57 2007 Error: 290010005 Host: darc Process: data_restore File: request_object.cxx Line: 556 Socket error: socket event from IG Error: 290009012 Socket event on 1, fd: 46. See long description for sample error message. Thu Feb 21 07:57:21 2008 Error: 290009012 Host: darc Process: recon_control File: IGControl.cpp Line: 1069 Socket event on 1, fd: 46. Error: 290009013 Not restarting IG from fd: 31 because we have reached the max number of consecutive resets. See LD for full Error Tue Jul 12 14:31:49 2005 290009013 4 darc recon_control recon_control_cb.cxx 460 Not restarting IG from fd: 31 because we have reached the max number of consecutive resets.
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Component
- 5159834-3 Jarrell VIG with Improved EMC Shielding, replaces 5159834-2 see long description for details 5159834-3 Jarrell VIG with Improved EMC Shielding, replaces 5159834-2 Part #: 5159834-3 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Jarrell VCT IG with VRAC2, Improved EMC Performance FRU: Yes Status: Orderable Per ECR 2042316, 212008898, 5/8/07. ** FE Assist ** FE must order software release 06MW29.7, p/n 5193989 , via FEMC/Coakly and perform LFC to support this FRU hardware if system is not currently running sw rls 06MW29.7, or more recent ( 5/15/07 212001261)
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required