Symptom

MOD drive is physically damaged and will not allow insert / removal of MOD cassettes.

Error message

Error: 10032 Media is not formatted or of unknown format See Long Description for the full error Sun Jun 19 22:29:28 2005 Error: 10032 Host: ct01 Process: arsjob File: ArchiveDevice.c Line: 4677 Media is not formatted or of unknown format

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GOC2 405L.8_H4.2M4 LightSpeed 4.X with GOC2 and MDAS 16 Gantry H4 Gantry Model # 2339985 and Global Operator Console 2 Model #'s 2366438 and 2366438-2(Hino) with HP8000 CPU (or newer) and Pegasus Recon ICE Box Product: LightSpeed16 405L.8_H4.2M4 CPU: GenuineIntel Intel(R) Xeon(TM) CPU 2.66GHz 4 2657 MHz i686 Processors Main memory size: 2048 Mbytes [root@presbyct2]# sg_map -i /dev/sg0 /dev/sda SEAGATE ST336753LW HPS2 /dev/sg1 /dev/sdb SEAGATE ST373453LW HPS1 /dev/sg2 /dev/sdc SONY SMO-F551-SD 1.05 /dev/sg3 /dev/scd0 MATSHITA DVD-RAM SW-9574S AZX8 /dev/sg4 /dev/scd1 HL-DT-ST DVD-ROM GDR8163B 0B15 [root@presbyct2]# df -h Filesystem Size Used Avail Use% Mounted on /dev/sda2 4.0G 1.7G 2.3G 41% / /dev/sda1 27M 12M 14M 44% /boot none 1008M 0 1007M 0% /dev/shm /dev/sda6 14G 3.5G 10G 25% /usr/g /dev/sda7 9.8G 753M 9.0G 8% /usr/g/sdc_image_pool /dev/sdb1 68G 28G 40G 41% /usr/g/sdc_image_pool2 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5167410 SCSI Tower with MOD and DVD drives for GOC series consoles (replaces 2368838 ) See Long Description for details Part #: 5167410 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: Peripheral Tower, dual-bay, MOD-DVDram FRU: Yes Status: Orderable Comments: ** FE ASSIST ** If ordering 5167410 (new) SCSI tower to replace 2368838 (old) SCSI tower for issues not related to the MOD, contact FMI Admin (Sahu, Shankar, sso 501375480 ) to order FMI 25408 (kit# 5317459) instead. FMI 25408 re-uses MOD, so if that is the failure, then continue with 5167410 order.(212001261 11/10/08) *** Second Opinion *** Reconfig should be tried prior to declaring DOA/FOA part (212001261 7/10/09) Per ECR 2030398, 8/2/06, MS11892.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.