Pular para o conteúdo principal
GE Healthcare · CTLightSpeedDiscovery24504323000006450003164000

LightSpeed 4.X - GRE

Symptom

"Scan hardware Stopped Scan" with Taxi data violation errors. "Scan hardware Stopped Scan" with Taxi data violation errors

Error message

Error: 245043 Scanning System Aborted Scanning. See LD for complete error Tue Feb 7 07:50:12 2006 Error: 245043 Host: ctscan1 Process: scanmgr File: scanmgr_events.cxx Line: 2115 Function: System Control : Scan Control Scan Type: Axial Scan: 0/0/0 Scan Seq Id: 1000 Scanning System Aborted Scanning

Error: 230000064 TAXI data violation detected by forward error correction See the long description for the full error code Sat Apr 17 20:26:43 2004 Error: 230000064 Host: darc Process: dataacq File: dip_thread.cxx Line: 411 Running diagnostic: Scan Data Path Functional Function: Data Acquisition : Scan Data Acquisition Scan Type: Axial Scan: 50003/5/1 Scan Seq Id: 164000 TAXI data violation detected by forward error correction. Interrupt bit: 4 Violation count: 1 Scan aborted.

Applies to

Models
LightSpeed, Discovery
Product
LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2318713-2 Fiber Optic cable Part is not a FRU at this time. Order the next higher FRU for your console type See LD for more Fiber Optic cable Part 2318713-2 is not a FRU at this time (10/30/05). Order 2293409 and use this part to get the system up. Part #: 2318713-2 Part Type: Not Available Part Weight: 0.00 LBR Repairable: Not Available Returnable: Not Available Hazardous Material: N Description: Not Available FRU: No Status: Non Orderable Part #: 2293409 Part Type: GP CONSUMABLES Part Weight: 0.40 LBR Repairable: No Returnable: No Hazardous Material: N Description: CABLE FIBER OPT FRU: Yes Status: Orderable Comments: FLS - CABLE FIBER OPT HARNESS ADDED VIA PCN 220275. USED ON WARP3. MS26561 DSM 8/25/01
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.