Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT2445402445182450111096187806178574178568

LightSpeed 1.X - GRE

Symptom

Scanner Hardware Reset Failed. See LD Scanner Hardware Reset Failed Hardware Failed to reset message appeared on monitor AFTER boot-up WITH Error:178568 Failed to establish socket connection to STC in 10 seconds and errors 244553 and 244554 and power up self test fails A, B, and 9 on STC CPU board The following errors were in the gesyslog Thu Jan 13 04:18:32 2005 Error: 178568 Suite: RP25 Host: SBC0 Process: fwmgr File: NBSSocket.m 1.9 Method: connectToTarget Line: 315 Failed to establish socket connection to STC in 10 seconds Thu Jan 13 04:18:32 2005 Error: 244553 Suite: RP25 Host: SBC0 Process: fwmgr File: HwSetUp.m 1.18 Method: hardReset Line: 429 The STC controller has failed to reset. Thu Jan 13 04:18:32 2005 Error: 244554 Suite: RP25 Host: SBC0 Process: fwmgr File: HwSetUp.m 1.18 Method: hardReset Line: 442 The reset failure of the OBC may be due to the reset failure of the STC. Also saw failures A, B, and 9 on the power-up test on the STC CPU .

Error message

Error: 244540 Scan Seq Id: 0 Immediate Request cannot be honored since the firmware is in the DOWN state See Long Description for complete error Sun Sep 26 08:36:46 2004 Error: 244540 4 000ctge04 fwmgr FwMgrEventHandler.c 1517 Running diagnostic: Scan Data Path Functional Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Immediate Request cannot be honored since the firmware is in the DOWN state

Error: 244518 Scan hardware reset failed See Long Description for complete error Sun Sep 26 08:30:10 2004 Error: 244518 ctge04 fwmgr HwManager.c 1536 Scanning hardware reset failed .

Error: 245011 The firmware wasn't available when a scan request was received See the LD for the full error code Error: 245011 The firmware wasn't available when a scan request was received 1096187806 2 1 Sun Sep 26 08:36:46 2004 245011 4 000ctge04 scanmgr validate_request.cxx 320 Running diagnostic: Scan Data Path Functional Function: System Control : Scan Control Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 The firmware wasn't available when a scan request was received. Rx has been told to reboot the firmware.

Error: 178574 The following controllers STC/ETC/OBC f

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 1.X - GRE H1_P_S4 Lightspeed 1.X - GRE Method of visual identification Large square gantry similar in appearance to a CTi gantry. Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Software
04MW13.7 for all LightSpeed systems with GRE/Xtream consoles (QXi, PLUS, ULTRA, LS16, PRO 16, RT)(1.X, 2.X, 3.X, 4.X, 5.X) 04MW13.7.(gan)_(tube)_(DAS)_(recon) for all LightSpeed systems with GRE/Xtream consoles (1.X, 2.X, 3.X, 4.X, 5.X) This is the new common Apps software for GRE/Xtream consoles. The GRE Common Applications software supports multiple hardware configurations via a single software set. The new software version string has the following format: <software_label>.<G>_<T>_<D>_<R> The "software label" format is: YYLWxx.ZZ Where: YY = Year. For example 04. L = Location. M - Milwaukee, H - Hino, B - Beijing. Wxx = Fiscal Week. For example, W15. ZZ = Revision number for each build/reroll. Refer to Table B for an explanation of the <G>_<T>_<D>_<R> format. Table B Common Applications Sotware Naming Convention G = GANTRY TYPE T = TUBE TYPE D = DAS TYPE R = RECON HW TYPE H1 = LightSpeed 1.X P = Performix S4 = SDAS4 P = Pegasus H2 = LightSpeed 2.X, 3.X, 4.X O = OEM M4 = MDAS4 G = GRE HP = LightSpeed 5.X H = Hercules M8 = MDAS8 WB = LightSpeed 5.x RT M16 = MDAS16 QX = HighSpeed QXi G16 = GDAS16 The variables G, T, D, and R are generated dynamically based on the type of hardware configuration this software is installed on. The information for the dynamic part of the string is obtained from the INFO files which are located in /usr/g/config directory. An example string based on the above format is as follows: 04MW13.7.H2_P_M4_G The break down is as follows: This is 2004 software release at Milwaukee in Fiscal Week 13 with a revision number "7". The hardware configuration is "H2" gantry with "Performix" tube, "MDAS4", and "GRE" recon hardware. CDL Link: Refer to: DIRECTION 2378261-100 GRE SOFTWARE INSTALLATION PROCEDURES
Area
CT

Full cause & step-by-step resolution