GE Healthcare · CTLightSpeed
LightSpeed RT16 & RT16 Xtra with GRE and GDAS16 gantry
Symptom
Applications software will not shutdown when desktop shutdown button is clicked on. Applications software will not shutdown when desktop shutdown button is clicked on
Applies to
- Models
- LightSpeed
- Product
- LightSpeed RT16 & RT16 Xtra and Vision WB_H_G16_G_GTL = RT16, Xtra WB_H_G16_G_GTL650 = Vision WB_H_G16_G_GTM = RT16, Xtra or Vision LightSpeed RT16 & RT16 Xtra with GRE and GDAS16 gantry Links to the Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Software
- 07BW08.3 For LightSpeed RT16 & RT16 Xtra ## Obsolete by FMI 22918 ## See LD for details 07BW08.3 For LightSpeed RT16 & RT16 Xtra - Issued by FMI 22915. Released 2nd Quarter 2007. LightSpeed RT16, Xtra Class C Software 07BW08.3 (Tab 182 distribution) 5224438 LightSpeed RT16, Xtra Class M Software 07BW08.3 (Tab 182 distribution) 5224439 Refer to Dir 5224323-100, Rev 2 which can be found on CT (LightSpeed) CDROM/DVD, or if online at: FCT Service Engineering Tech Pubs Website, or Common Document Library. Always use the most appropriate documentation for your system. The following was taken from Dir 5224323-100, Rev 2 for your preferences See FMI document for complete information. FMI22915 kit (5224323) consists of the following: Item Part Number Description Quantity 1. 5224323-100 FMI22915 Instruction 1 2. 5199147 LS RT16 and Xtra Software Collector for 07BW08.3 Software 1 A. 5199653 Software Collector 1 1. 5199245 Application 07BW08.3 Release 1 2. 5215358 Linux Operating System 6.2.9 1 3. 5219992 Serial over LAN CD 1 B. 46-303268P1 Rating Plate 1 C. 5224951 WB Service Pack A Software For 07BW08.3 Release on LS RT16, Xtra 1 D. 5220681 Service Pack B Software For 07BW08.3 1 Release on LS RT16, Xtra; BS Select Series 3. 5219970-100 LS RT16, Xtra Application Tips and Work-Arounds_English 1 4. 5219970-247 LS RT16, Xtra Application Tips and Work- 1 Arounds Multi-Language CD
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required