Pular para o conteúdo principal
GE Healthcare · CTLightSpeed244555178575178568

LightSpeed 2.X (Plus, QX/i) - Octane

Symptom

Scanning HW reset fails to reset OBCR

Error message

Error: 244555 The OBC controller has failed to reset See Long Description for the full error Error: 244555 Host: ct01_sbc Process: fwmgr File: HwSetUp.c Line: 724 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 The OBC controller has failed to reset.

Error: 178575 Failed to R-HARD the following controllers: STC OBC ETC. See long description for full error Fri Oct 8 07:49:44 2004 178575 2 ct03 fwmgr HwSetUp.c 837 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Failed to R-HARD the following controllers: STC OBC ETC.

Error: 178568 Failed to establish socket connection to OBCR in 10 seconds. Sun Nov 27 14:18:24 2005 Error: 178568 Suite: SBCT Host: SBC0 Process: fwmgr File: NBSSocket.m 1.9 Method: connectToTarget Line: 315 Failed to establish socket connection to ETC in 10 seconds ------------------------------------------------------------- Wed Sep 8 06:20:02 2004 Error: 178568 Host: cmls Process: fwmgr File: NBSSocket.c Line: 388 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Failed to establish socket connection to OBCR in 10 seconds. ------------------------

Applies to

Models
LightSpeed
Product
LightSpeed 2.X (Plus, QX/i) - Octane Always S-DAS 2.4.2_H2.4M5 LightSpeed 2.X (Plus, QX/i) - Octane (Pegasus or Rhapsode IG) - Always S-DAS Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 2.4.2_H2.4M5 FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 FEATURES/OPTIONS STD/OPTION Cat # Digital Laser Camera I/F B7700PD Smart Helical (s/w only) Standard B7530TK Fast Recon (s/w only) Option B7700JF BMD (requires B7710MP or MN) Option B7710MK Smart Prep Option B7530SP Smart Score ? ? 3D Analysis (Requires D3D) Option B7540TD Navigator (s/w only) Option B7540N DentaScan (s/w only) Option B7540SD Direct 3D (Requires B7710SA) Option B7810SS HIS/RIS Option (h/w and s/w) Option B7500PN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2377227 UTP Transceiver or Media Convertor used on OC with 1X - 4X Gantry - replaces 2148559 10BASE-2 TO UTP CONVERTER See Long Description for Details. 2377227 UTP Transceiver or Media Covnerter for GRE GOC3 Consoles with HP-8000 CPU and older style Gantries LS1X - 4X Part #: 2377227 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: UTP TRANSCEIVER - BNC TO RJ45 With I.T.E PWR SPPLY 24000027 FRU: Yes Status: Orderable Part #: 2148559 Part Type: GP CONSUMABLES Part Weight: 1.36 LBR Repairable: No Returnable: No Hazardous Material: N Description: 10BASE-2 TO UTP CONVERTER FRU: Yes Status: Orderable Comments: ***OBSOLETE PART Case # 11345706*** This part has been identified as technically obsolete and no short term solution has been identified. FE must identify and order a higher or lower level spare as required. Contact your support engineering, On-line Engineering or Parts ID team should you require support. 212012173 7-24-49 ** FE ASSIST ** For P/N 2148559 - Please order the newer media converter 2377227 and external terminator 2154019. The 2154019 50 ohm external terminator is required.(212001261 10/2/07) I re addded my note 7/28/08 as someone removed it
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.