Pular para o conteúdo principal
GE Healthcare · CTLightSpeed23001301524504730000230005024230006530000000

LightSpeed 4.X - GOC2

Symptom

Multiple recon retries and suspended images during an exam. See long description for details Multiple recon retries and suspended images during an exam. To get the images to recon you would have to resume the queue or reboot the system for the images to recon.

Error message

Error: 230013015 xx DAS views of 100 retrieved from the scan database failed checksums See Long Description for great info on this error Problem: The Following errors all point to a scsi Scan data disk issue. Using the gesys_Ct01.log and iceConsole.log is a must, together both logs point to the Scan data disk chain. Check DIP stats if stats are clean the das data is good. The iceConsole log is a little difficult to read and hard to correlate to specific dates and times. --------------gesys_CT01.log--------------- Sun Jan 13 20:33:00 2002 Error: 245047 Host: gect88 Process: scanmgr File: scan_sequence.cxx Line: 1377 Function: System Control : Scan Control Scan Type: Helical Scan: 0/0/0 Scan Seq Id: 30000 Unable to allocate scan files. Sun Jan 13 20:33:05 2002 Error: 230013015 Host: ice Process: rebin File: RebinMDASData.cxx Line: 332 Function: Image Generation : Data Restore Job Type: Prospective Helical Scan: 31/2/1 Image: 31/2/56 Recon Seq Id: 173 xx DAS views of 100 retrieved from the scan database failed checksums <-- DAS Das fails checksum from Scan data disk!!!! Sun Jan 13 20:33:05 2002 Error: 230005024 Host: ice-sm-gate Process: sdcctrl File: sdcctrl_events.cxx Line: 358 imag

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GOC2 405L.8_H4.2M4 LightSpeed 4.X with GOC2 and MDAS 16 Gantry H4 Gantry Model # 2339985 and Global Operator Console 2 Model #'s 2366438 and 2366438-2(Hino) with HP8000 CPU (or newer) and Pegasus Recon ICE Box Product: LightSpeed16 405L.8_H4.2M4 CPU: GenuineIntel Intel(R) Xeon(TM) CPU 2.66GHz 4 2657 MHz i686 Processors Main memory size: 2048 Mbytes [root@presbyct2]# sg_map -i /dev/sg0 /dev/sda SEAGATE ST336753LW HPS2 /dev/sg1 /dev/sdb SEAGATE ST373453LW HPS1 /dev/sg2 /dev/sdc SONY SMO-F551-SD 1.05 /dev/sg3 /dev/scd0 MATSHITA DVD-RAM SW-9574S AZX8 /dev/sg4 /dev/scd1 HL-DT-ST DVD-ROM GDR8163B 0B15 [root@presbyct2]# df -h Filesystem Size Used Avail Use% Mounted on /dev/sda2 4.0G 1.7G 2.3G 41% / /dev/sda1 27M 12M 14M 44% /boot none 1008M 0 1007M 0% /dev/shm /dev/sda6 14G 3.5G 10G 25% /usr/g /dev/sda7 9.8G 753M 9.0G 8% /usr/g/sdc_image_pool /dev/sdb1 68G 28G 40G 41% /usr/g/sdc_image_pool2 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.