GE Healthcare · Nuclear MedicineInfinia100008553

During boot of Infinia NWSERVER process is using all the CPU time

Symptom

During boot of Infinia NWSERVER process is using all the CPU time When Infinia server boot we can see under: Windows Task Manager -> processes -> nwserver.exe take almost 100% of CPU time and do not decrease. To check if it's your case, when Infinia Server boot type: Ctrl+Alt+delete , choose: Task Manager , select tab: Processes , click: CPU, this will sort by CPU usage. You will see processes nwserver.exe always on top. Contributor 100008553 B.Certo October 2, 2008

Error message

SM_SERVER_CRASH_ERR:.:Server: (NWSERVER) has crashed:.::.: Server: (NWSERVER) has crashed:.:Trace:{1} under syslog.csv If you check file syslog.csv under problem report we can FATAL error: SM_SERVER_CRASH_ERR:.:Server: (NWSERVER) has crashed:.::.: Server: (NWSERVER) has crashed:.:Trace:{1} If you check file [email protected] under log directory: 17:18:05 SMS: smsManager::SetLMsStatus- LM - LMTEST, Stat=0 17:18:05 REGISTER_WD returned :: GDU status=0 17:18:05 SMS: SmsSignalHandler::handle_timeout- WD reply from LM (GDU) is OK. 17:18:05 SMS: smsManager::SetLMsStatus- LM - GDU, Stat=0 17:18:17 ERROR=Error in SystemMonitor d:\tacq\su\src\cpp\services\comm\communit\scommunitex\scommunitex.cpp(1278): Failed sending opcode REGISTER_WD to server NWSERVER, 5 tries exhausted 17:18:17 SMS: SmsSignalHandler::handle_timeout- Error in reply from LM (NWSERVER), Activate Action- 2 17:18:17 17:18:17 ********************** SERVERS STATISTICS ******************** 17:18:17 NAME FAILURES (crashes) 17:18:17 CTReconServer 0 17:18:17 CTServer 0 17:18:17 DBAcc 0 17:18:17 FEserver 0 17:18:17 GDU 0 17:18:17 GantryServer 0 17:18:17 LMTEST 0 17:18:17 MouseLimit 0 17:18:17 NWSERVER 2 17:18:17 Reframer 0 17

Applies to

Models
Infinia
Product
Infinia I Infinia I Acquisition System Infinia HE4 Infinia II Hawkeye 4 & Infinia II GP (Infinia running Windows XP software) Infinia II Infinia II Acquisition System
Software
Infinia All Infinia All. Including Infina 1 and Infinia II. Infinia All. Including Infina 1 and Infinia II.
Component
*DICOM DICOM DICOM
Area
Nuclear Medicine

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare Nuclear Medicine issue are available to FSELIB subscribers — re-authored, web-first, searchable.