GE Healthcare · Nuclear MedicineInfinia100008553
During boot of Infinia NWSERVER process is using all the CPU time
Symptom
During boot of Infinia NWSERVER process is using all the CPU time When Infinia server boot we can see under: Windows Task Manager -> processes -> nwserver.exe take almost 100% of CPU time and do not decrease. To check if it's your case, when Infinia Server boot type: Ctrl+Alt+delete , choose: Task Manager , select tab: Processes , click: CPU, this will sort by CPU usage. You will see processes nwserver.exe always on top. Contributor 100008553 B.Certo October 2, 2008
Error message
SM_SERVER_CRASH_ERR:.:Server: (NWSERVER) has crashed:.::.: Server: (NWSERVER) has crashed:.:Trace:{1} under syslog.csv If you check file syslog.csv under problem report we can FATAL error: SM_SERVER_CRASH_ERR:.:Server: (NWSERVER) has crashed:.::.: Server: (NWSERVER) has crashed:.:Trace:{1} If you check file [email protected] under log directory: 17:18:05 SMS: smsManager::SetLMsStatus- LM - LMTEST, Stat=0 17:18:05 REGISTER_WD returned :: GDU status=0 17:18:05 SMS: SmsSignalHandler::handle_timeout- WD reply from LM (GDU) is OK. 17:18:05 SMS: smsManager::SetLMsStatus- LM - GDU, Stat=0 17:18:17 ERROR=Error in SystemMonitor d:\tacq\su\src\cpp\services\comm\communit\scommunitex\scommunitex.cpp(1278): Failed sending opcode REGISTER_WD to server NWSERVER, 5 tries exhausted 17:18:17 SMS: SmsSignalHandler::handle_timeout- Error in reply from LM (NWSERVER), Activate Action- 2 17:18:17 17:18:17 ********************** SERVERS STATISTICS ******************** 17:18:17 NAME FAILURES (crashes) 17:18:17 CTReconServer 0 17:18:17 CTServer 0 17:18:17 DBAcc 0 17:18:17 FEserver 0 17:18:17 GDU 0 17:18:17 GantryServer 0 17:18:17 LMTEST 0 17:18:17 MouseLimit 0 17:18:17 NWSERVER 2 17:18:17 Reframer 0 17Applies to
- Models
- Infinia
- Product
- Infinia I Infinia I Acquisition System Infinia HE4 Infinia II Hawkeye 4 & Infinia II GP (Infinia running Windows XP software) Infinia II Infinia II Acquisition System
- Software
- Infinia All Infinia All. Including Infina 1 and Infinia II. Infinia All. Including Infina 1 and Infinia II.
- Component
- *DICOM DICOM DICOM
- Area
- Nuclear Medicine
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare Nuclear Medicine issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare Nuclear Medicine issues
- Service Note 43783_1 Feb/08 Infinia/Ventri/Xeleris 1.1 & 2
- After software upgrade FMI40786, the Infinia application does not automatically start on boot up
- I need the product data sheet on catalog E4500AK x-ray in use warning light interface control panel for Infinia
- Infinia Does not recognize collimator is mounted to detector, not reading collimator ID correctly.
- Where do I obtain the Non-Linear Correction tables for a VG/Infinia Hawkeye system if can't find the CD that came with the DAS assembly?
- Hawkeye CT scan Ring (Circular) artifacts in X-ray images appearing in the same radius in all of the images.