Symptom

Recon shuts down with no access scan database for scan

Error message

Error: 230009026 Could not access scan database for scan.... Thu Feb 2 17:08:47 2006 Error: 230009026 Host: uvct1 Process: queuemgr File: qmgr_list_mgmt.c Line: 619 Function: Image Generation : SBC Processing Job Type: Prospective Helical Scan: 21828/2/1 Image: 21828/3/29 Could not access scan database for scan

Error: 230011000 Scan data disk handling not initialized. A successful initialization of the Scan Disk I/O subsystem must occur prior to writing or accessing scan data See LD for the full error Tue Aug 23 08:03:24 2005 Error: 230011000 Host: ice Process: sdiomgr File: sdiomgr.cxx Line: 2315 initRawDataDisk failed Scan data disk handling not initialized. A successful initialization of the Scan Disk I/O subsystem must occur prior to writing or accessing scan data

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GOC2 405L.8_H4.2M4 LightSpeed 4.X with GOC2 and MDAS 16 Gantry H4 Gantry Model # 2339985 and Global Operator Console 2 Model #'s 2366438 and 2366438-2(Hino) with HP8000 CPU (or newer) and Pegasus Recon ICE Box Product: LightSpeed16 405L.8_H4.2M4 CPU: GenuineIntel Intel(R) Xeon(TM) CPU 2.66GHz 4 2657 MHz i686 Processors Main memory size: 2048 Mbytes [root@presbyct2]# sg_map -i /dev/sg0 /dev/sda SEAGATE ST336753LW HPS2 /dev/sg1 /dev/sdb SEAGATE ST373453LW HPS1 /dev/sg2 /dev/sdc SONY SMO-F551-SD 1.05 /dev/sg3 /dev/scd0 MATSHITA DVD-RAM SW-9574S AZX8 /dev/sg4 /dev/scd1 HL-DT-ST DVD-ROM GDR8163B 0B15 [root@presbyct2]# df -h Filesystem Size Used Avail Use% Mounted on /dev/sda2 4.0G 1.7G 2.3G 41% / /dev/sda1 27M 12M 14M 44% /boot none 1008M 0 1007M 0% /dev/shm /dev/sda6 14G 3.5G 10G 25% /usr/g /dev/sda7 9.8G 753M 9.0G 8% /usr/g/sdc_image_pool /dev/sdb1 68G 28G 40G 41% /usr/g/sdc_image_pool2 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2339285 GOC1/GOC2 SCSI Control Card See Long Description Part #: 2339285 Part Type: GP CONSUMABLES Part Weight: -19999.80 LBR Repairable: No Returnable: No Hazardous Material: N Description: PMC SCSI CONTROLLER CARD FRU: Yes Status: Orderable Comments: FLS - PMC SCSI CONTROLLER CARD. ADDED PER H16 PROGRAM. ADDED BY MS20544 ON 7/19/02. 8/7/02- C2R COLLECTION PER P REILLY. MS12464 11/11/02 PROCUREMENT MFG/CT 7/11/03 - ROUTED TO SWAP. MS12575
Area
CT

Full cause & step-by-step resolution