GE Healthcare · CTHiSpeed2540743079055422
RP sub-system failed its self test.
Symptom
RP sub-system failed its self test. Data collection may be impaired. Reconstructed images may contain artifacts. RP sub-system failed its self test. Data collection may be impaired. Reconstructed images may contain artifacts
Error message
Error: 254074 IG0 DSP0: Backprojection error See long Description for full error Sat Nov 17 06:47:51 2007 Error: 254074 Suite: CT01 Host: SBC Process: Diag Exec File: IGBp.c Method: IG0 BST Line: 1485 Function: Diagnostics No System Function Reported Test Name: SBC IG0 Datapath Unique ID: 30790 Iteration: 1 Exc. Level: Soft Time: 0 Error Type: Firmware IG0 DSP0: Backprojection error. The PRAU Image Memory scan complete value was incorrect. Probable cause: IG board Expected: 0x5020 Actual: 0x0 Mon Sep 15 07:50:42 2008 Error: 254074 Suite: CT99 Host: SBC Process: Diag Exec File: IGBp.c Method: IG0 BST Line: 1485 Function: Diagnostics No System Function Reported Test Name: SBC IG0 Datapath Unique ID: 55422 Iteration: 1 Exc. Level: Soft Time: 0 Error Type: Firmware IG0 DSP0: Backprojection error. The PRAU Image Memory scan complete value was incorrect. Probable cause: IG board Expected: 0x5020 Actual: 0x0
Applies to
- Models
- HiSpeed
- Product
- HiSpeed CTi 6.X HiSpeed CTi 6.X with Octane and HSA Gantry HiSpeed CTi 6.X with Octane and HSA Gantry Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required