GE Healthcare · CTHiSpeed2538011601617600118540161931521

Scanner Hardware Stopped Scan

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 253801 View Length Error. The size of the views received were incorrect See Long Description for complete error Mon Dec 6 12:12:06 2004 Error: 253801 Suite: RHCT Host: FEP Process: CoordinateMode File: fepin.c 1.61 Method: FEPIN_err_handler Line: 2355 Function: Data Acquisition : Scan Data Acquisition Scan: 16016/2/5 Type: Helical Scan Seq: 176001 FEP DSP 0: View Length Error. The size of the views received were incorrect (HDW status: 0x4006b7) (Data Type: 0x1). View Sequence Number: 6537. Possible causes: Slip Ring buffer terminator boards, SCOM board, or FEP board

Error: 185401 slip-ring view length errors 1 See Long Description for full error Host: SBC0 Proc: fwmgr Error: 185401 Exception Level: Pri/Soft SARQ statistics: total views 6151, retried subpackets 2 slip-ring view length errors 1 DAS view length error 0, DAS parity error 0 Taxi Link violations -61931521

Applies to

Models
HiSpeed
Product
HiSpeed CT/i 4.X HiSpeed CT/i 4.x with Zeus HSA Gantry and SGI Indigo Console HiSpeed CT/i with:HSA GANTRY CT/I (RHAPSODE) CONSOLE (Indigo Computer) Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.