Symptom
Table would not move in or out. Table would not move in or out
Error message
Error: 260115245 MTCB cradle encoder delta data out of range. See long description for full sample error. Thu Aug 7 08:34:27 2008 Error: 260115245 Host: tgp Process: tgp subsys File: mtcb cradle monitor.c Line: 285 Thu Aug 7 08:34:27 2008 Host : Table Sub-System Ermes # : 260115245 Exception Class : Secondary Severity : Pri/Soft File : mtcb cradle monitor.c Line# : 285 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 MTCB cradle encoder delta data out of range. Possible Encoder, cabling or MTCB issue. cradle current position = 23755 cradle previous position = 17710 Pacsci drive command position = -96 position offset = -396 previous position offset = -396 Error: 260115236 MTCB Cradle Drive did not stop when expected Tue May 30 16:57:09 2006 Error: 260115236 Host: tgp Process: tgp_subsys File: mtcb_cradle_monitor.c Line: 487 Tue May 30 16:57:08 2006 Host : Table Sub-System Ermes # : 260115236 Exception Class : Soft Severity : Pri/Soft File : mtcb_cradle_monitor.c Line# : 487 Function : No System Function Reported Scan Type : None/Unknown Scan: 17521/3/0 MTCB Cradle Drive did not stop when expected. Target position = 4294967295 Current Position = 132453 Sta
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required