Pular para o conteúdo principal
GE Healthcare · CTLightSpeed2600031582600010235072326000235850011260003117

LightSpeed 4.X - GRE

Symptom

"Scan hardware Stopped Scan" with collimator errors. "Scan hardware Stopped Scan" with collimator errors

Error message

Error: 260003158 The Filter Home Flag Has Moved Since The Last Reference See long description for full error Fri Feb 17 17:56:41 2006 Error: 260003158 Host: HSC Process: CCB Firmware File: filter_home.c Line: 193 Function: XRAY GENERATION : Xray Collimation and Filtration Scan: 8650/2/0 Type: None/Unknown Exception Level: Sec/Soft Time: 17:56:41:390 Log Series: 564 The Filter Home Flag Has Moved Since The Last Reference. Possible Causes: Filter Home Switch, Filter Encoder, Filter Encoder Set Screw. Difference: 360 degrees

Do Not Use - use one of the other two Error: 260001023 See Long Description for complete error NOTE: The error message code included in this error is variable. You will need to refer to the log to find the cause of this error. Tue Dec 7 07:23:22 2004 Error: 260001023 Host: OBC Process: HOP File: hop_erlog.c Line: 367 Function: No System Function Reported Scan: 50723/1/1 Type: Axial Exception Level: Pri/Most Severe Time: 07:23:22:910 Log Series: 37 The DCB has notified the system of an error condition which prevents scanning. See log for further information: Error message code: 260002358 Notification level: 5 Sequence number: 4 Fri May 11 06:44:04 2007 Error: 2600

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5140001-3 (replaces 5140001-2) Helios 3 collimator with notched filler bar and rating plate (used on 1.X-4.X) See Long Description for more info Part #: 5140001-3 Part Type: GP REPAIRABLES Part Weight: 115.00 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: helios 3 collimator with notched filler bar and rating plate FRU: Yes Status: Orderable Comments: ** FE Assist ** 1.x 2.x 3.x 4.x started with Ultras added minor quality improvements to the dash 2 and is backwards compatible (12/6/06 212001261) Part #: 5140001-2 Part Type: GP REPAIRABLES Part Weight: 0.00 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: helios 2 collimator with notched filler bar and rating plate FRU: Yes Status: Orderable Comments: This part is replacement for 2214768-2, please do not use stock of 2214768-2. ms13452 050516 Can use the 5140001-3 if no Stock of the -2 (9/26/05 212001261)
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.