Pular para o conteúdo principal
GE Healthcare · CTHiSpeed25382216016180001035331960012033968

"Scan hardware Stopped Scan" with DAS power errors.

Symptom

"Scan hardware Stopped Scan" with DAS power errors. "Scan hardware Stopped Scan" with DAS power errors

Error message

Error: 253822 HPDAS Power supply auxiliary channels out of tolerance See Long Description for complete error Error: 253822 HPDAS Power supply auxiliary channels out of tolerance Mon Dec 6 12:12:30 2004 Error: 253822 Suite: RHCT Host: FEP Process: CoordinateMode File: fepin.c 1.61 Method: FEPIN_dma_done Line: 1822 Function: Data Acquisition : Scan Data Acquisition Scan: 16016/2/5 Type: Helical Scan Seq: 180001 FEP DSP 1:HPDAS Power supply auxiliary channels out of tolerance View #2 Values in volts. 753: +5: 0.035331 <==== +6 Analog Supply Too Low 754: -5: 4.960012 755: +5: 5.033968 756: +15: 26.452396 <==== +15 Analog Supply Too High 757: -15: 15.327940 Note: Pay attention to the specific Power Supply values. Different values relate to different causes.

Error: 253801 View Length Error. The size of the views received were incorrect See Long Description for complete error Mon Dec 6 12:12:06 2004 Error: 253801 Suite: RHCT Host: FEP Process: CoordinateMode File: fepin.c 1.61 Method: FEPIN_err_handler Line: 2355 Function: Data Acquisition : Scan Data Acquisition Scan: 16016/2/5 Type: Helical Scan Seq: 176001 FEP DSP 0: View Length Error. The size of the views received were incorrect (HD

Applies to

Models
HiSpeed
Product
HiSpeed CTi 6.X HiSpeed CTi 6.X with Octane and HSA Gantry HiSpeed CTi 6.X with Octane and HSA Gantry Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2298032 HSA & CT/I EMC SLIPRING CONVERSION KIT 2298032 HSA & CT/I EMC SLIPRING CONVERSION KIT 2298032 CONTAINS EVERY THING NEEDED TO UPGRADE FROM AN ETC RING TO A SCHLLEIF RING USED ON A HSA OR CTI EMC GANTRY. DO NOT MIX ETC AND SCHLIEF RING BRUSHES ON THE SAME RING 2298032 HSA & CT/I EMC SLIPRING CONVERSION KIT 2298032 CONTAINS EVERY THING NEEDED TO UPGRADE FROM A ETC RING TO A SCHLLEIFRING USED ON A HSA OR CTI EMC GANTRY. DO NOT MIX ETC AND SCHLIEFRING BRUSHES ON THE SAME RING 2298034 FOR HSA & CT/I NON-EMC SLIPRING KITS ONLY 2298034 FOR HSA & CT/I NON-EMC SLIPRING KITS ONLY DO NOT MIX ETC AND SCHLIEFRING BRUSHES ON THE SAME RING 2298034 FOR HSA & CT/I NON-EMC SLIPRING KITS ONLY DO NOT MIX ETC AND SCHLIEFRING BRUSHES ON THE SAME RING
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.