Infinia gantry is not finishing collimator exchange hung with no radial motion.

Symptom

Infinia gantry is not finishing collimator exchange hung with no radial motion. No errors are generated. No motions are available.

Applies to

Models
Infinia
Product
Infinia I Infinia I Acquisition System Infinia HE4 Infinia II Hawkeye 4 & Infinia II GP (Infinia running Windows XP software) Infinia II Infinia II Acquisition System
Component
CBL001315 (5366338) Collimator Exchange switch assembly H1 Infinia CBL001315 (5366338) Collimator Exchange switch assembly H1 Infinia The new switch 5366338 will have activation force of 625g to 800g compare to 114g to 350g of the old switch. This make the new switch more resistive to false activation due to detector vibrations CBL001308 (5366344) Collimator Exchange switch assembly H2 Infinia CBL001308 (5366344) Collimator Exchange switch assembly H2 Infinia The new switch 5366344 will have activation force of 625g to 800g compare to 114g to 350g of the old switch. This make the new switch more resistive to false activation due to detector vibrations *Gantry Gantry Gantry
Area
Nuclear Medicine

Full cause & step-by-step resolution