Symptom

"Scanner Hardware Stopped Scan" DURING SMARTVIEW prep WITH Interventional Hand Held Control (IHHC) failed to enable Scanner Hardware Stopped Scan DURING SMARTVIEW prep WITH Interventional Hand Held Control (IHHC) failed to enable See the log description for the full error code Mon Sep 19 10:35:34 2005 Error: 182700 Suite: CT01 Host: ETC Hand Held Control Process: Diag Exec File hc_e xec.c 1.38 No Method Line: 595 Function: Diagnostics OPERATOR INPUT/OUTPUT : Tableside Operator Interface Test Name: Unique ID: 58982 Iteration: 1 Exception Level: Hard Time: 18945 Error Type: Firmware Interventional Hand Held Control (IHHC) failed to enable. Verify IHHC is connected to scanner. If IHHC is not connected to scanner, connect IHHC to scanner. If IHHC is connected to scanner, disconnect IHHC from scanner and reconnect IHHC to scanner. Thu Aug 14 12:26:27 2008 Error: 182701 Host: Process: gr:RECEIVE:firmware File: hc_wdog.c 1.7 2/16/99 Line: 255 Function: No System Function Reported Scan: 25796/3/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 12:26:26:920 Log Series: 418 Interventional Hand Held Control (IHHC) Watchdog Timeout. The IHHC has been disabled. HW/SW Error: Probable hardware, possible software, problem

Error message

Error: 200109123 Received BAD SmartStep Start Status From Firmware. Please check connection for the Hand Held Control Unit. See Long Description for complete error. Error: 200109123 Received BAD SmartStep Start Status From Firmware. Please check connection for the Hand Held Control Unit. Fri Sep 19 09:49:46 2008 Error: 200109123 Host: smhct3 Process: scanRx File: uifIpcMgrScanRx.c Line: 2198 Function: Data Acquisition : OC Processing Received BAD SmartStep Start Status From Firmware, status = -1, seq_num = 306. P lease check connection for the Hand Held Control Unit.

Error: 182700 Interventional Hand Held Control (IHHC) failed to enable See the long description for the full error code Mon Sep 19 10:35:34 2005 Error: 182700 Suite: CT01 Host: ETC Hand Held Control Process: Diag Exec File hc_e xec.c 1.38 No Method Line: 595 Function: Diagnostics OPERATOR INPUT/OUTPUT : Tableside Operator Interface Test Name: Unique ID: 58982 Iteration: 1 Exception Level: Hard Time: 18945 Error Type: Firmware Interventional Hand Held Control (IHHC) failed to enable. Verify IHHC is connected to scanner. If IHHC is not connected to scanner, connect IHHC to scanner. If IHHC is connected to scanner, disco

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - GRE Always 4 or 8 slice MDAS. H2_P_M4 H2_P_M8 LightSpeed 3.X - GRE Method of visual identification Small rounded gantry similar in appearance to LS 2.X and 4.X. Always has an MDAS4 or MDAS8 Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da Always has an MDAS4 or MDAS8 List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2199947 CT SONATA HAND HELD CONTROLLER CT SONATA HAND HELD CONTROLLER Part number 2199947 CT SONATA HAND HELD CONTROLLER Part number 2199947
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.