Pular para o conteúdo principal
GE Healthcare · CTLightSpeed1019116896

LightSpeed 4.X - GRE

Symptom

MOD Save or Restore functions fail. MOD Save or Restore functions fail Troubleshooting There is no media in the drive message appears in a attention box during a MOD Save or Restore of with Error: 10191 Failed save exam. posted in the Message Log. The fllowing pop-up message appears on screen. Problem is consistent If a different disk is used then the error occurs again when trying to save, restore, or label

Error message

Error: 10191 Failed save exam. Mon Nov 27 03:45:02 2006 Error: 10191 Host: emc1 Process: arserver File: arscommand.c Line: 2304 Failed save exam. Paused 16896

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - GRE See Long Description for detailed information - H2_P_M16 Product Name: LightSpeed 4.X with GRE Product Photographs GE HEALTHCARE Method of remote or software identification - See attachment below for a complete list of software commands and typical results 1. check_config 2 . swhwinfo “swhwinfo” will have the following switches: · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console types: GOC3 (GRE) GOC4 (GRE) List of software the product can run: 07MW11.10 SP2.0, SP3.0, SP3.1 FAQ's Table weight limit - 180 kg (400 lbs) normal operation - 205 kg (450 lbs) maximum allowed with decreased accuracy Tube type - D3186T D3186T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2324697-2 SCSI CABLE, VHD68 TO CENT 50. see LD for a drawing Part #: 2324697-2 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: SCSI CABLE, VHD68 TO CENT 50, 6 FT FRU: Yes Status: Orderable Comments: FLS - SCSI CABLE, VHD68 TO CENT 50, LINUX. NEED THIS CABLE IF LINUX SYSTEM WITH DIGITAL DASM CAN HITLINE MFG MS27898 VENDOR IS NORTECH, MN USA PDR ON FILE IM05 REQUESTED
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.