HiSpeed CTi 6.X with Octane and HSA Gantry
Symptom
Exam/Series/Image did not show up in the Patient List Select (image browser). See Long Description for Details Exam/Series/Image did not show up in the Patient List Select (image browser). The customer did a refresh of the browser and the images were then shown in the list. Selecting the images and trying to display them resulted in a message “No Valid Selection”. User scanned a series and the images were generated and displayed in the autoview view port. The issue is that the Exam/Series/Image did not show up in the Patient List Select (image browser). The customer did a refresh of the browser and the images were then shown in the list. Selecting the images and trying to display them resulted in a message “No Valid Selection”. Images did get put in the database but were not accessible by the customer. Reboot did not fix this problem
Error message
Error: 200000436 Image not installed because image available event flagged image as not being available. Fri May 27 09:40:15 2005 Error: 200000436 Host: ct2 Process: dms File: InstallImage.cxx Line: 1113 Image 4243/1/1 not installed because image available event flagged image as not being available.
Applies to
- Models
- HiSpeed
- Product
- HiSpeed CTi 6.X HiSpeed CTi 6.X with Octane and HSA Gantry HiSpeed CTi 6.X with Octane and HSA Gantry Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 46-320144G1 STEALTH-X CPU, 32 MB, SBC board see long description Part #: 46-320144G1 Part Type: GP REPAIRABLES Part Weight: 3.64 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: STEALTH-X CPU, 32 MB FRU: Yes Status: Orderable Comments: OBSOLETE PART case 4971762 This part has been identified as obsolete and no short-term solution has been identified FE must identify and order a higher or lower level spare part as required Contact your support engineer on-line engineer or Parts ID team should you require support
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.