Pular para o conteúdo principal
GE Healthcare · CTLightSpeed2601163075100151000

LightSpeed 5.X (80/100kW)

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 260116307 Could not send message to 101 or 201 from 101 or 201 (any combination) see long description for error details. Error: 260116307 Could not send message to 101 or 201 from 101 or 201. See examples below: Fri Aug 15 15:01:01 2008 Error: 260116307 Host: tgp Process: tgp_subsys File: lan_watchdog.c Line: 513 Fri Aug 15 15:00:48 2008 Host : Table Gantry Processor Ermes # : 260116307 Exception Class : Secondary Severity : Sec/Soft File : lan_watchdog.c Line# : 513 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Could not send message to 201 from 201. ( TGP from TGP) Event Attempting to send 51001 System Error - 0 Sun Jul 5 13:49:14 2009 Error: 260116307 Host: tgp Process: tgp subsys File: lan watchdog.c Line: 568 Sun Jul 5 14:00:47 2009 Host : Table Gantry Processor Ermes # : 260116307 Exception Class : Secondary Severity : Sec/Soft File : lan watchdog.c Line# : 568 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 Could not send message to 101 from 201. (ORP from TGP) Event Attempting to send 51000 Tue Dec 15 14:28:25 2009 Error: 260116307 Host: orp Process: orp subsys File: lan watchdog.c Line: 568 Tue Dec 15 14:24:2

Applies to

Models
LightSpeed
Product
LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5126615 ASTEC Digital Power Supply Asm for GDAS 4/8/16 See long description for details 5126615 ASTEC Digital Power Supply Asm for GDAS 4/8/16 Part #: 5126615 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: ASTEC Digital PS Assembly FRU: Yes Status: Orderable Comments: Added per RT / GDAS RSPL
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues