Pular para o conteúdo principal
GE Healthcare · CTLightSpeed

LightSpeed RT16 & RT16 Xtra with GRE and GDAS16 gantry

Symptom

Table squeaks creaks scrapes is noisy when driven in or out of the gantry. Table squeaks creaks scrapes is noisy when driven is or out of the gantry.

Applies to

Models
LightSpeed
Product
LightSpeed RT16 & RT16 Xtra and Vision WB_H_G16_G_GTL = RT16, Xtra WB_H_G16_G_GTL650 = Vision WB_H_G16_G_GTM = RT16, Xtra or Vision LightSpeed RT16 & RT16 Xtra with GRE and GDAS16 gantry Links to the Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5168472-2 New style IMS bearing replacement kit for GT1700 or GT2000 tables. See long description for details Part #: 5168472-2 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: IMS DRIVE REPLACEMENT SET Positioning GT FRU: Yes Status: Orderable There is a new Kit for the VCT Table IMS. The old kit is no longer available. The old Kit part number will automatically change up to the number for the new kit: 5168472-2 . FE's are confused by the parts in the Kit and there are no installation instructions. FMI 26809 will be released with instructions: The Kit contains only one Bearing, but the Table has 2 Bearings. This is correct . The new Bearing is larger and as a result only one Bearing is required now. There are no instructions in the Kit on how to install all the parts. Use the hyperlink below to get instructions on how to install the Kit. http://jptyoiis01.ap.health.ge.com:8000/ctdoc/FMI/19/26809.pdf
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues