GE Healthcare · CTLightSpeed244584178570260001008260001085
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane
Symptom
Customer gets Error message 'DAS gain calibration has not been run since converter boards were changed' while running fast cals. Customer gets Error message 'DAS gain calibration has not been run since convertor boards were changed' while running fast cals.
Error message
Error: 244584 Collimator/DAS reset failed See Long Description for complete error Error: 244584 Host: qx-ct Process: fwmgr File: HwManager.c Line: 1536 Collimator/DAS reset failed. Error: 178570 Coll/Das controller reports failure in initializing tasks See Long Description for complete error Mon Mar 28 03:39:48 2005 Error: 178570 Host: ct04 Process: fwmgr File: HwSetUp.c Line: 3178 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 OBC_SS timeout: Coll/Das controller reports failure in initializing tasks. Do Not Use - use one of the other Error: 260001008 The OBC failed to connect with the ? (DCB and/or the CCB) sub-system in the allotted time Open Long Description to see error example Mon Mar 28 03:39:48 2005 Error: 260001008 Host: OBC Process: HOP File: hop_node_init.c Line: 120 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 03:39:46:170 Log Series: 31 The OBC failed to connect with the DCB (could be the CCB) sub-system in the allotted time. Check to see if subsystem is alive. Timeout value: 40000mS Error: 260001085 Communication error: Message not sent to the DCB. Communications with t
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 46-170021P75 15A 250 V DAS power fuse. See Long description for details P/N 46-170021P75 15A 250 V DAS power fuse Part #: 46-170021P75 Part Type: GP CONSUMABLES Part Weight: 0.01 LBR Repairable: No Returnable: No Hazardous Material: N Description: 015.000A 0250V 3AB SLOW UL FRU: Yes Status: Orderable
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.