Symptom
Customer sees "Detector Temperature Is High. Image quality may be affected". Customer sees Detector Temperature Is High. Image quality may be affected
Error message
Error: 260007416 Detector Temperature Is High. Image quality may be affected See Long Description Tue Sep 6 09:07:45 2005 Error: 260007416 Host: orp Process: orp_subsys File: dtemptst.c Line: 364 Tue Sep 6 09:07:44 2005 Host : Das Ermes # : 260007416 Exception Class : Soft Severity : Sec/Soft File : dtemptst.c Line# : 364 Function : No System Function Reported Scan Type : Scout Scan: 372/1/1 Detector Temperature Is High. Image quality may be affected. Current Reading:40.41 Deg C A/D Counts:20053 Error: 260002471 Data Acquisition System Temperature is nearing Upper Limit See LD for complete error Thu Feb 16 22:19:58 2006 Host Das Proc: DCB Firmware Error: 260002471 File: ctemptst.c Line: 784 Function: DATA ACQUISTION : Scan Data Acquistion Scan: scan 50024/1/1 Type: Axial Exception: Pri/Soft Time: 22:19:58:760 Log Series: 579 WARNING : Data Acquistion System Temperature is nearing Upper Limit
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5340738 (replaces 2359114) HPOWER GANTRY FAN AND HARNESS used on 5.x gantry See Long Description for details This fan for BrightSpeed, HP100, HP60 and RT Systems. Part #: 5340738 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: GANTRY FAN AND HARNESS FRU: Yes Status: Orderable Comments: Per ECR 2091297, 212008898, 10/26/09. Part #: 2359114 <-- All GPRS stock is gone and this part is now obsolete Part Type: GP CONSUMABLES Part Weight: 8.82 LBR Repairable: No Returnable: No Hazardous Material: N Description: HPOWER GANTRY FAN AND HARNESS FRU: Yes Status: Orderable Comments: FLS - HPOWER GANTRY FAN AND HARNESS.ADDED PER THE HPOWER RSPL. EXPECTED M3 FOR THIS PROGRAM: FW19, 2003. MS20544 2/28/03.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.