Pular para o conteúdo principal
GE Healthcare · CTHiSpeed25382468001000000771880

HiSpeed CTi 6.X with Octane and HSA Gantry

Symptom

"Scan hardware Stopped Scan" with DAS power errors. "Scan hardware Stopped Scan" with DAS power errors

Error message

Error: 253824 EDAS Power supply auxiliary channels out of tolerence. See Long Description for Details. Fri Jul 18 07:52:12 2008 Error: 253824 Suite: ONC1 Host: FEP Process: CoordinateMode File: fepin.c 1.61 Method: FEPIN dma done Line: 1905 Function: Data Acquisition : Scan Data Acquisition Scan: 2928/1/4 Type: Axial Scan Seq: 68001 FEP DSP 1:EDAS Power supply auxiliary channels out of tolerence. View #1 Values in volts. 759: +5:0.000000 <<=== Reading Low 756:-15:15.771880 757:+15:0.000000 <<=== Reading Low

Applies to

Models
HiSpeed
Product
HiSpeed CTi 6.X HiSpeed CTi 6.X with Octane and HSA Gantry HiSpeed CTi 6.X with Octane and HSA Gantry Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
46-183027P129 ADC Low Noise EDAS See Long Description for Details Part #: 46-183027P129 Part Type: GP REPAIRABLES Part Weight: 1.36 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: ADC FRU: Yes Status: Orderable
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required