GE Healthcare · CTLightSpeed26000107150039
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane
Symptom
Scan aborts WITH Communications with the CCB sub-system has not been established. Scan aborts WITH Communications with the CCB sub-system has not been established
Error message
Error: 260001071 The CCB has reset unexpectedly. The OBC has successfully reconnected to sub-system. See LD for complete error Error: 260001071 The CCB has reset unexpectedly. The OBC has successfully reconnected to sub-system. Fri Feb 10 16:23:44 2006 Error: 260001071 Host: OBC Process: HOP File: hop_node_init.c Line: 347 Function: No System Function Reported Scan: 50039/6/0 Type: None/Unknown Exception Level: Pri/Soft Time: 16:23:44:745 Log Series: 130 The CCB has reset unexpectedly. The OBC has successfully reconnected to sub -system. Reason for reset as reported by CPU: External signal source. CPU reset register: 0x80 Node status: 0x382 State flags: 0x9800682
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2159578-3 collimator control board used on 1X/2X/3X/4X/5X See Long description for details Part #: 2159578-3 Part Type: GP CONSUMABLES Part Weight: 2.18 LBR Repairable: No Returnable: No Hazardous Material: N Description: COLLIMATOR CONTROL BOARD FRU: Yes Status: Orderable Comments: Not Available
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required