Pular para o conteúdo principal
GE Healthcare · CTLightSpeed2600040255022919127260007840

LightSpeed 3.X (Ultra, Plus, QX/i) - Octane

Symptom

Unable to scan with a message on screen that says: Collimator/DAS reset failed.

Error message

Error: 260004025 DAS / Collimator firmware object: Lost communications with node applications. See Long Description for full error Thu Jan 13 06:58:23 2005 Error: 260004025 Host: OBC Process: Collimator File: col_cf_in_mail.c Line: 442 Function: No System Function Reported Scan: 50229/1/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 06:58:23:235 Log Series: 47 Collimator firmware object: Lost communications with node applications. Thu Dec 31 11:45:03 2009 Error:260004025 Host: OBC Process: DAS File: das cf in mail.c Line: 430 Function: No System Function Reported Scan: 19127/2/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 11:45:03:305 Log Series: 451 DAS firmware object: Lost communications with node applications.

Error: 260007840 DCB-to-DHCB Communications Failed See long description for full error Tue Nov 16 11:59:22 2004 Error: 260007840 Host: DAS Process: DCB Firmware File: dhcbcomm.c Line: 459 Function: DATA ACQUISITION : Scan Data Acquisition Scan: 4801/2/0 Type: None/Unknown Exception Level: Pri/Soft Time: 11:59:22:975 Log Series: 595 DCB-to-DHCB Communications Failed - Exhausted Comm retries. The DCB Sent The following Command to the DHCB. DHCB T

Applies to

Models
LightSpeed
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2225212-4 PWR SUPPLY ASM, 24V F-CASE See Long Description Part #: 2225212-4 Part Type: GP CONSUMABLES Part Weight: 19.18 LBR Repairable: No Returnable: No Hazardous Material: N Description: PWR SUPPLY ASM, 24V F-CASE FRU: Yes Status: Orderable Parts Offered as: Current : 2225212-4 Harvested : Not Available Exchange : Not Available Comments: FLS - PWR SUPPLY ASM, 24V F-CASE ADDED VIA PCN 220176. USED ON HELIOS 2. MS26561 DSM 6/26/01
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues