GE Healthcare · CTLightSpeedVCT2600010235072326000235850011260003117260007391
LightSpeed 3.X (Ultra, Plus, QX/i) - GRE
Symptom
"Scan hardware Stopped Scan" with DAS power errors. "Scan hardware Stopped Scan" with DAS power errors
Error message
Do Not Use - use one of the other two Error: 260001023 See Long Description for complete error NOTE: The error message code included in this error is variable. You will need to refer to the log to find the cause of this error. Tue Dec 7 07:23:22 2004 Error: 260001023 Host: OBC Process: HOP File: hop_erlog.c Line: 367 Function: No System Function Reported Scan: 50723/1/1 Type: Axial Exception Level: Pri/Most Severe Time: 07:23:22:910 Log Series: 37 The DCB has notified the system of an error condition which prevents scanning. See log for further information: Error message code: 260002358 Notification level: 5 Sequence number: 4 Fri May 11 06:44:04 2007 Error: 260001023 Host: OBC Process: HOP File: hop_erlog.c Line: 374 Function: No System Function Reported Scan: 50011/1/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 06:44:05:145 Log Series: 42 The CCB has notified the system of an error condition which prevents scanning. See log for further information:^C Error message code: 260003117 Notification level: 5 Sequence number: 4 Error: 260007391 The Data Acquisition System Has Detected An Error In The -/+12V analog DAS Power Supply See LD for more details Tue Aug 8 07:49:0
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 3.X (Ultra, Plus, QX/i) - GRE Always 4 or 8 slice MDAS. H2_P_M4 H2_P_M8 LightSpeed 3.X - GRE Method of visual identification Small rounded gantry similar in appearance to LS 2.X and 4.X. Always has an MDAS4 or MDAS8 Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da Always has an MDAS4 or MDAS8 List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5191271 MDAS 16 Fuse Holder Upgrade Kit. Supercedes part number 46-170042P28. See Long Description for more information. Part #: 5191271 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: MDAS 16 Fuse Holder Upgrade Kit FRU: Yes Status: Orderable Comments: Per ECR 2036788, 212008898, 11/30/2006. MDAS Fuse Holder Replacement Kit Contents: Part #: 46-170042P28 Part Type: GP CONSUMABLES Part Weight: 2.18 LBR Repairable: No Returnable: No Hazardous Material: N Description: 30A, 250V PANEL-MOUNT FUSEHOLDER. FRU: Yes Status: Orderable Comments: FLS - FUSE HOLDER PART OF FMI KIT ** FE Assist ** IF FUSE HOLDER P/N 46-170042P28 FAILED, PLEASE ORDER MDAS FUSE FIX ON FAILURE KIT 5191271 (12/12/06 212001261)
- Area
- CT
Full cause & step-by-step resolution